DwyerOmega DRC-8900
DwyerOmega

DRC-8900 Series isolated DIN rail signal conditioners for TC, RTD, V and mA

Signal conditioners & barriers

Manufacturer
DwyerOmega
Model
DRC-8900
Instrument type
Signal conditioners & barriers
Inputs
thermocouple J, K; Pt RTD 100/500/1000 Ω; DC voltage; DC current
Outputs
0–10 V or 4–20 mA
Accuracy
±0.1 % FS (V/I); ±2 °C (TC); ±0.7 °C (RTD)
Response time
10 ms (V/I); 1.5 s (TC/RTD)
Isolation
three-way, 700 V DC (power, input, output)
CMRR
85 dB
Load resistance
0 to 500 Ω
Open-circuit voltage
13 V
Power supply
22–26 V DC unregulated
Operating temperature
0 to 55 °C
Storage temperature
−40 to 80 °C
Housing
plastic UL-94V-1, 22.5 × 110 × 77 mm
Weight
142 g
Mounting
35 mm DIN rail
Request a quote
Review enquiry

The DRC-8900 Series consists of DIN rail mounted signal conditioners for process monitoring. These isolated transmitters take signals from thermocouples, resistance thermometers, DC voltage or DC current sources and convert them into standardised output signals suitable for data acquisition systems, controllers and recorders. The articles DRC-8901 to DRC-8908 differ in input type, input range and output signal.

Inputs are configurable by model: type J and K thermocouples, platinum RTDs with 100, 500 or 1000 Ω, DC voltage in various mV and V ranges, and DC current in mA ranges. The output is either 0–10 V or 4–20 mA. The signals are linearised, and thermocouple models include cold junction compensation, so the output is proportional to temperature without further processing in the receiving system.

Stated accuracy is ±0.1 % of full scale for voltage and current inputs, ±2 °C for thermocouples and ±0.7 °C for RTDs. Response time is 10 ms for voltage and current models and 1.5 s for temperature models. The common mode rejection ratio is 85 dB. The current output drives loads from 0 to 500 Ω, with an open-circuit voltage of 13 V.

All models provide three-way isolation rated at 700 V DC between power supply, input and output circuits. This separates the field signal from the measuring system and helps prevent ground loops. The conditioners require an unregulated DC supply of 22 to 26 V.

The slim housing is made of UL-94V-1 plastic with an integrated DIN rail foot and measures 22.5 mm in width, 110 mm in height and 77 mm in length; the weight is 142 g. The modules snap onto a 35 mm DIN rail such as RAIL-35-1 or RAIL-35-2. Permitted operating temperature is 0 to 55 °C, storage temperature −40 to 80 °C.

The series is intended for process monitoring and data acquisition systems. Please state the sensor type or input range and the required output signal in your enquiry, so that the correct DRC-8900 article can be selected.

Part numbers · 21

DRC-8901, DRC-8902, DRC-8903, DRC-8904, DRC-8905, DRC-8906, DRC-8907, DRC-8908, DRC-8911, DRC-8912, DRC-8913, DRC-8914, DRC-8915, DRC-8916, DRC-8917, DRC-8918, DRC-8920, DRC-8921, DRC-8940, DRC-8941, DRC-8942

Industry applications

Selection notes

List input types and the destination of each output. Match panel cutout, supply and communication options before choosing a replacement.

For an installed instrument, copy the complete ordering code and its suffixes from the nameplate. Identify the part that will remain in service and the part that needs to be replaced. A photograph of the connection face or terminal markings helps resolve differences between versions that share the same housing.

For a new installation, describe what the instrument must measure or control and where the reading will be used. Give the normal operating conditions separately from short peaks. Include units with every numerical value. A drawing with the available space, orientation and access for removal gives the supplier a practical starting point.

Add the quantity for each position and identify any accessories that belong with it. Where a particular drawing, material certificate or calibration record is required, mention it with the line item. Submit the list with the destination country so that the response can cover the requested configuration and delivery.

Related instruments