PACE6000-ADC
Calibration

Temperature calibrators
The 4180 series are portable precision infrared calibrators for point IR thermometers and thermal imagers. Each unit provides a large, temperature-controlled flat target that the instrument under test views, so that its reading can be compared with a known radiance temperature. The calibrators are used in the laboratory and in the field.
Target temperature is controlled by a digital controller with a precision platinum RTD sensor. The 4180 uses FET-driven Peltier modules and can therefore both cool and heat; the 4181 uses a heater driven by a solid-state relay (triac). The target surface has a nominal emissivity of 0.95 (0.928 for the 4180ETRS) and is radiometrically calibrated, and emissivity compensation can be set between 0.9 and 1.0 to match the setting of the thermometer being tested. An accredited radiometric calibration is included.
Three models cover different needs. The 4180 works from −15 to 120 °C with display accuracy from ±0.40 °C at −15 °C to ±0.55 °C at 120 °C. The 4181 covers 35 to 500 °C with accuracy from ±0.35 °C at 35 °C to ±1.60 °C at 500 °C. The 4180ETRS is a dedicated version for fever-screening systems, working from 33 to 40 °C with ±0.2 °C accuracy and stability of ±0.006 °C over one minute; it includes a USB converter. Target uniformity and stability are specified for 5-inch and 2-inch diameter areas.
Heating and cooling cycles are rapid: the 4180 heats from −15 to 120 °C in 15 minutes and cools from 23 to −15 °C in 20 minutes; the 4181 heats from 35 to 500 °C in 45 minutes. The calibrators store eight set-points and offer scan rate control. Remote operation is possible over RS-232 with SCPI-1994 syntax.
The 152.4 mm (6 in) target is large enough for instruments with wide fields of view. The units weigh between 9.1 and 9.5 kg. They conform to IEC 61010-1 and IEC 61326-1, and the radiometric calibration follows ASTM E1256-95. Accessories include the 4180-APRT aperture, the 4180-CASE carrying case and the 4180-DCAS case with wheels.
Applications include calibration of infrared thermometers and thermal imagers, fever-screening systems, laboratory temperature reference and HVAC testing. Please state the temperature range you need to cover and the type of instruments to be calibrated in your enquiry.
4180, 4181, 4180ETRS
Identify the instrument under test and the points to reproduce. State the required uncertainty, reference conditions and certificate format.
For an installed instrument, copy the complete ordering code and its suffixes from the nameplate. Identify the part that will remain in service and the part that needs to be replaced. A photograph of the connection face or terminal markings helps resolve differences between versions that share the same housing.
For a new installation, describe what the instrument must measure or control and where the reading will be used. Give the normal operating conditions separately from short peaks. Include units with every numerical value. A drawing with the available space, orientation and access for removal gives the supplier a practical starting point.
Add the quantity for each position and identify any accessories that belong with it. Where a particular drawing, material certificate or calibration record is required, mention it with the line item. Submit the list with the destination country so that the response can cover the requested configuration and delivery.