EDT 101
Temperature

Pyrometers & thermal cameras
The 64 MAX is a handheld infrared thermometer for electrical, industrial, HVAC and solar work. It measures surface temperatures without contact from a distance, with a 20:1 distance-to-spot ratio, and stores readings in an internal memory of 99 data points.
The instrument detects the infrared energy radiated by the target in the 8 to 14 µm spectral band and converts it into a temperature reading. A single-point laser marks where the thermometer is aimed. Emissivity can be adjusted from 0.10 to 1.00 to suit the surface being measured. Results are shown on a backlit LCD with 0.1 °C resolution, and the response time is below 500 ms for 95 % of the reading.
The measuring range is −30 °C to 600 °C. Accuracy is ±1.0 °C or ±1.0 % of reading, whichever is greater, at a calibration ambient of 23 °C ±2 °C; between −10 °C and 0 °C it is ±2.0 °C and between −30 °C and −10 °C ±3.0 °C. Repeatability is ±0.5 % of reading or ±0.5 °C, whichever is greater.
Several functions support routine checks. Trigger lock allows continuous measurement, and auto capture records up to 99 measurements at a set time and interval for unattended monitoring; readings are stamped with date and time. Hi/Lo alarms signal temperatures outside a set band, and the unit can display °C or °F. A flashlight is built in.
The thermometer is rated IP54 to IEC 60529 and tested for a 3 m drop. It is designed to IEC 61010-1, pollution degree 2, and the laser is Class 2 to IEC 60825-1 (650 nm, < 1 mW). One AA battery powers it for 30 hours with laser and backlight on. It weighs 255 g, measures 175 × 85 × 75 mm and operates from 0 °C to 50 °C up to 2,000 m altitude.
Related models with lower specifications are the 62 MAX+ (−30 to 650 °C, 12:1) and the 62 MAX (−30 to 500 °C, 10:1). Kits and a tripod mounting accessory are available.
Listed applications are electrical, industrial, HVAC and solar work, temperature monitoring and industrial maintenance.
Please let us know the required quantity and whether a kit or tripod accessory is needed.
64 MAX IR Thermometer
Match the sensing element, insertion length and wiring to the existing input. Separate the process temperature from the ambient temperature at the connection head.
For an installed instrument, copy the complete ordering code and its suffixes from the nameplate. Identify the part that will remain in service and the part that needs to be replaced. A photograph of the connection face or terminal markings helps resolve differences between versions that share the same housing.
For a new installation, describe what the instrument must measure or control and where the reading will be used. Give the normal operating conditions separately from short peaks. Include units with every numerical value. A drawing with the available space, orientation and access for removal gives the supplier a practical starting point.
Add the quantity for each position and identify any accessories that belong with it. Where a particular drawing, material certificate or calibration record is required, mention it with the line item. Submit the list with the destination country so that the response can cover the requested configuration and delivery.