Optris CTratio 2M
Optris

CTratio 2M fibre-optic two-colour ratio pyrometer, 250–3000 °C

Pyrometers & thermal cameras

Manufacturer
Optris
Model
CTratio 2M
Instrument type
Pyrometers & thermal cameras
Measuring principle
two-colour (ratio), also one-colour mode
Temperature range
250–3000 °C (versions 2ML, 2MH, 2MH1)
Spectral range
1.45–1.75 µm
Response time
1 ms
Optics
D:S 38:1 or 100:1; standard or close focus; focus 150 mm to infinity
Accuracy
±0.5 % of reading + 2 °C
Repeatability
±0.3 % of reading
NETD
100 mK
Sighting, fibre cable
green laser; fibre 3 m, 8 m or 15 m
Ambient, sensing head
−20 to 200 °C
Outputs, I/O
2 × 0/4–20 mA, 0–5/0–10 V, relay, 3 programmable I/Os, 0–10 V input
Interfaces
USB, RS-232, RS-485, Ethernet; Modbus RTU/TCP, PROFINET, EtherNet/IP
Power
8–30 VDC or 5 V USB, max. 5 W
Protection
IP65, NEMA 4
Size, weight
head Ø 19 × 76 mm, 210 g; electronics 122 × 70 × 27 mm, 420 g
Request a quote
Review enquiry

The CTratio 2M is a rugged fibre-optic two-colour (ratio) pyrometer for precise non-contact temperature measurement in high-temperature industrial environments, from 250 to 3000 °C. Application areas named include metalworking, semiconductor manufacturing, advanced materials, glass processing and plastic manufacturing.

The two-colour principle compares the radiation at two nearby infrared wavelengths. Because the temperature is derived from the ratio, the measurement stays accurate even when the emissivity varies or the sighting window becomes contaminated. A Smart Ratio Mode (SRM) is included, and the instrument can also measure in one-colour mode. The optical head is connected to the electronics by fibre-optic cable, and a green laser helps align the measuring spot; focus is adjustable from 150 mm to infinity.

The spectral range is 1.45–1.75 µm and the response time 1 ms. Accuracy is ±0.5 % of reading + 2 °C, repeatability ±0.3 % of reading and thermal sensitivity (NETD) 100 mK. Three versions are offered: CTratio 2ML for 250–1000 °C (one-colour) or 275–1000 °C (two-colour) with D:S 38:1; CTratio 2MH for 375–1500 °C or 400–1500 °C with D:S 100:1; and CTratio 2MH1 for 500–3000 °C or 550–3000 °C with D:S 100:1. Standard or close-focus optics and fibre cable lengths of 3 m, 8 m or 15 m are available.

The electronics provide two 0/4–20 mA outputs, 0–5 V and 0–10 V, a relay, three programmable I/Os and a 0–10 V input. Digital interfaces are USB, RS-232, RS-485 and Ethernet with Modbus RTU, Modbus TCP, PROFINET and EtherNet/IP. Power is 8–30 VDC or 5 V via USB, at most 5 W. Configuration is done with PC software or a mobile app.

The sensing head, Ø 19 × 76 mm and 210 g, operates at −20 to 200 °C; the electronics measure 122 × 70 × 27 mm and weigh 420 g. Protection is IP65/NEMA 4, at 10–95 % non-condensing humidity. A high-temperature option and isolation options are available, and a mounting nut and bracket are included. The instrument is tested to IEC 60068-2-6, -2-27 and -2-64, the laser complies with EN 60825-1, and it is CE marked and RoHS compliant.

Please state the temperature range, optics, fibre cable length and required interface when enquiring.

Part numbers · 3

OPTCTRF2ML, OPTCTRF2MH, OPTCTRF2H1

Industry applications

Selection notes

Match the sensing element, insertion length and wiring to the existing input. Separate the process temperature from the ambient temperature at the connection head.

For an installed instrument, copy the complete ordering code and its suffixes from the nameplate. Identify the part that will remain in service and the part that needs to be replaced. A photograph of the connection face or terminal markings helps resolve differences between versions that share the same housing.

For a new installation, describe what the instrument must measure or control and where the reading will be used. Give the normal operating conditions separately from short peaks. Include units with every numerical value. A drawing with the available space, orientation and access for removal gives the supplier a practical starting point.

Add the quantity for each position and identify any accessories that belong with it. Where a particular drawing, material certificate or calibration record is required, mention it with the line item. Submit the list with the destination country so that the response can cover the requested configuration and delivery.

Related instruments