Fluke 709H
Fluke

709H precision mA loop calibrator with HART communication

Multifunction process calibrators

Manufacturer
Fluke
Model
709H
Instrument type
Multifunction process calibrators
Functions
mA source, mA simulate, mA read, mA read with loop power, volts read
Ranges
0 to 24 mA; 0 to 30 V DC
Resolution
1 µA (mA); 1 mV (V)
Accuracy
0.01 % ±2 counts, all ranges (at 23 ±5 °C)
Loop power
24 V DC at 20 mA
Load capability
Up to 1200 Ω
HART resistor
250 Ω, selectable
Auto ramp / step time
5 to 300 s
Data logging
9800 records
Display
128 × 64 pixel graphic LCD with backlight
Power
6 × AAA alkaline, approx. 40 h
Protection
240 V AC overvoltage, 28 mA DC overload; IP40
Operating temperature
−10 to 55 °C
Dimensions / weight
15 × 9 × 3 cm; 0.3 kg
Request a quote
Review enquiry

The 709H is a precision mA loop calibrator for installation, calibration and troubleshooting of field transmitters, valves and other control system components in process plants. In addition to the analogue functions of a loop calibrator it contains a HART modem, so it can communicate with HART-enabled instruments.

The calibrator sources, simulates and measures DC current from 0 to 24 mA and measures DC voltage up to 30 V. The operating modes are mA source, mA simulate (two-wire transmitter simulation), mA read, mA read with loop power, and volts read. A built-in 24 V DC loop supply powers a transmitter during testing, with a loop compliance voltage of 24 V at 20 mA and load capability up to 1200 Ω. With the HART modem the user can read device information, perform diagnostics and trim the calibration of HART transmitters; a selectable internal 250 Ω resistor provides the loop resistance needed for HART communication. Automatic step and ramp functions run with times from 5 to 300 seconds.

Accuracy is 0.01 % ±2 counts on all ranges at 23 ±5 °C, with a resolution of 1 µA on the mA range and 1 mV on the voltage range. Temperature stability is 20 ppm of full scale per °C outside the reference band. The input is protected against overvoltage up to 240 V AC and overload current up to 28 mA DC.

Operation uses a rotary encoder knob for quick setting of values and a backlit 128 × 64 pixel graphic display. The instrument stores up to 9800 data records, which can be transferred over USB. It runs on six AAA alkaline cells for about 40 hours, with auto shutdown selectable from 1 to 30 minutes or disabled. It measures 15 × 9 × 3 cm and weighs 0.3 kg; the operating temperature range is −10 to 55 °C, protection is IP40 and EMC conforms to EN 61326 Annex A.

Delivery includes a NIST-traceable calibration certificate with data, batteries, test leads and manual. Track software with cable is available, as well as a German user interface. Related models are the 709 without HART and the 710, which adds test functions for HART smart control valves.

Please specify whether the 709H, 709 or 710 is required, the optional software and the quantity when you enquire.

Part numbers · 1

709H

Industry applications

Selection notes

Identify the instrument under test and the points to reproduce. State the required uncertainty, reference conditions and certificate format.

For an installed instrument, copy the complete ordering code and its suffixes from the nameplate. Identify the part that will remain in service and the part that needs to be replaced. A photograph of the connection face or terminal markings helps resolve differences between versions that share the same housing.

For a new installation, describe what the instrument must measure or control and where the reading will be used. Give the normal operating conditions separately from short peaks. Include units with every numerical value. A drawing with the available space, orientation and access for removal gives the supplier a practical starting point.

Add the quantity for each position and identify any accessories that belong with it. Where a particular drawing, material certificate or calibration record is required, mention it with the line item. Submit the list with the destination country so that the response can cover the requested configuration and delivery.

Related instruments

Find equipment

DEVICEEU