Fluke 715
Fluke

715 volt/mA loop calibrator

Multifunction process calibrators

Manufacturer
Fluke
Model
715
Instrument type
Multifunction process calibrators
Current measure / source
0 to 24 mA (−25 % to 125 %), resolution 0.001 mA
Current accuracy
±(0.01 % + 2 counts) (firmware 2.0+)
Voltage measurement
0 to 200 mV / 0 to 20 V (firmware 2.0+)
Voltage source
up to 200 mV, 20 V or 25 V (firmware 2.0+); 100 mV / 10 V earlier
Voltage resolution
0.01 mV / 0.001 V
Voltage accuracy
±(0.015 % + 2) mV; ±(0.01 % + 2) V
Loop power
24 V DC nominal
HART resistor
250 Ω, built in
Output functions
step and linear ramp; configurable zero and span
Display
5-digit LCD
Battery
one 9 V alkaline
Operating temperature
−10 to 55 °C
Size
187 × 87 × 32 mm
Weight
approx. 330 g
Shock and vibration
random 2 g, 5–500 Hz; 1 m drop
Request a quote
Review enquiry

The 715 is a compact, full-featured volt/mA loop calibrator for measuring and sourcing process signals. It is built to be rugged and lightweight, with a simple push-button interface and a clear five-digit LCD, and it covers the routine tasks of checking and calibrating current loops, transmitters and PLC inputs and outputs.

In measure mode the 715 reads loop current in the 0–20 mA and 4–20 mA ranges and voltage outputs from PLCs and transmitters. In source mode it sources or simulates 4–20 mA loop current and sources DC voltage. A built-in 24 V loop supply powers a transmitter while the calibrator measures its current at the same time, and a 250 Ω resistor for HART communication is integrated. Step and linear ramp functions and configurable zero and span settings help with linearity checks.

Current is measured and sourced from 0 to 24 mA (−25 % to 125 % of a 4–20 mA span) with 0.001 mA resolution and an accuracy of 0.01 % + 2 counts with firmware 2.0 and later (0.02 % + 2 counts on earlier firmware). Voltage is measured in 200 mV and 20 V ranges with firmware 2.0 and later, with accuracy of 0.015 % + 2 counts in millivolts and 0.01 % + 2 counts in volts and resolution of 0.01 mV and 0.001 V. The voltage source covers 100 mV or 10 V on earlier firmware and up to 200 mV, 20 V or 25 V on newer firmware, which also adds fuseless input protection.

Power comes from a single 9 V alkaline battery. The calibrator operates from −10 to 55 °C, is stored from −40 to 60 °C and can be used up to 3,000 m altitude. It measures 187 × 87 × 32 mm and weighs about 330 g. It withstands random vibration of 2 g from 5 to 500 Hz and a 1 m drop, and it is EMI tolerant, dust and splash resistant. Safety is to CSA C22.2 No. 1010.1, and EMC to EN 50082-1 and EN 55022 Class B. Optional Ex versions are available for hazardous areas.

Accessories include a protective holster with test lead storage, test leads, alligator clips, a quick reference guide and an NIST-traceable calibration report with data.

Typical uses are process calibration, loop checks and the calibration, repair and maintenance of current loops. Please indicate whether you need the standard or an Ex version and which accessories should be included in your enquiry.

Part numbers · 1

715

Industry applications

Selection notes

Identify the instrument under test and the points to reproduce. State the required uncertainty, reference conditions and certificate format.

For an installed instrument, copy the complete ordering code and its suffixes from the nameplate. Identify the part that will remain in service and the part that needs to be replaced. A photograph of the connection face or terminal markings helps resolve differences between versions that share the same housing.

For a new installation, describe what the instrument must measure or control and where the reading will be used. Give the normal operating conditions separately from short peaks. Include units with every numerical value. A drawing with the available space, orientation and access for removal gives the supplier a practical starting point.

Add the quantity for each position and identify any accessories that belong with it. Where a particular drawing, material certificate or calibration record is required, mention it with the line item. Submit the list with the destination country so that the response can cover the requested configuration and delivery.

Related instruments

Find equipment

DEVICEEU