PACE6000-ADC
Calibration

Pressure calibrators & pumps
The 718Ex is an intrinsically safe, self-contained pressure calibrator for work in explosion-hazardous areas. It combines pressure measurement, current measurement and a built-in hand pump in one portable unit, so that pressure transmitters and pressure switches can be checked on site without a separate pump or loop meter.
Pressure is generated with the integrated pressure/vacuum hand pump, which has a fine-adjust vernier and a bleed valve for setting test points precisely. The internal sensor measures the applied pressure, while a separate input measures DC current up to 24 mA. Both values are shown at the same time, which allows a P/I transmitter to be calibrated by comparing the applied pressure with its output current. A pressure switch test function and Min/Max hold are also provided.
Three versions with internal sensors are available: 30G from −12 to 30 psi, 100G from −12 to 100 psi and 300G from −12 to 300 psi (−82.7 kPa to 2068 kPa). Resolution is 0.001 psi for the 30G and 0.01 psi for the other two ranges. Accuracy of the internal sensor is 0.05 % of full span. For higher pressures the calibrator connects to external 700PEx pressure modules covering 2.5 mbar to 700 bar, with accuracy to 0.025 % of full span. Overpressure limits are 3 × FS for the 30G, 2 × FS for the 100G and 375 psi (25 bar) for the 300G. Current is measured with 0.02 % of reading ± 2 counts and 0.001 mA resolution.
The instrument is approved to ATEX II 1G Ex ia IIC T4 and carries CSA certification for intrinsically safe use in Class I, Division 1, Groups A–D, T4; EMC conformity follows EN 61326. It is intended for non-corrosive gases. Power comes from a 9 V carbon-zinc battery. The operating temperature range is −10 to 55 °C, storage −40 to 60 °C and the maximum operating altitude 3,000 m. With its holster the calibrator measures 216 × 94 × 66 mm and weighs about 992 g.
Accessories and options include the 700PEx pressure modules, external pumps such as the 750PEx, a protective holster, TL75 test leads and AC72 test clips.
Typical tasks are calibration of P/I transmitters, pressure switch testing and maintenance of pressure systems in potentially explosive atmospheres. Please state the required pressure range and whether external modules are needed when you send your enquiry.
2096310, 2096322, 3023328, 718Ex
Identify the instrument under test and the points to reproduce. State the required uncertainty, reference conditions and certificate format.
For an installed instrument, copy the complete ordering code and its suffixes from the nameplate. Identify the part that will remain in service and the part that needs to be replaced. A photograph of the connection face or terminal markings helps resolve differences between versions that share the same housing.
For a new installation, describe what the instrument must measure or control and where the reading will be used. Give the normal operating conditions separately from short peaks. Include units with every numerical value. A drawing with the available space, orientation and access for removal gives the supplier a practical starting point.
Add the quantity for each position and identify any accessories that belong with it. Where a particular drawing, material certificate or calibration record is required, mention it with the line item. Submit the list with the destination country so that the response can cover the requested configuration and delivery.