PACE6000-ADC
Calibration

Multifunction process calibrators
The 726 is a handheld precision multifunction calibrator for the process industry. It measures and sources the main process parameters, voltage, current, resistance, frequency and temperature, and together with the 750P series pressure modules it also calibrates pressure instruments. It is used to calibrate transmitters, to test sensors and valves and for flowmeter testing.
The calibrator has two separate channels, so it can measure and source signals at the same time. For example, it can source mA while measuring pressure, which allows valves and transmitters to be tested. A built-in 24 V loop supply (max. 22 mA) powers a two-wire transmitter during the test, and an integrated 250 Ω resistor allows HART communication to be used in the loop. The 726 interprets calibration results and stores up to eight of them in memory. An integrated pressure switch test function is also provided.
DC voltage is measured from 0 to 30.000 V and sourced from 0 to 20 V. Current can be sourced or simulated from 0 to 24.000 mA. Resistance is measured and sourced from 0.0 to 4000.0 Ω with ±0.015 % accuracy, and frequency from 1.0 Hz to 15.00 kHz with 0.05 % + 0.1 Hz.
Pressure is measured with external pressure modules that use silicon piezoresistive sensors. The 750P series offers a wide range of modules, for example the 750PA6, 750PA7, 750PA8, 750PA9, 750PA27, 750PV3 and 750PV4; 700P series modules can be used as well. To generate the pressure, the 700PTP-1 pneumatic test pump and the 700HTP-2 hydraulic test pump are available.
The calibrator measures 200 × 96 × 47 mm, weighs 650 g and runs on four AA alkaline batteries. It operates from −10 to +55 °C and up to 3000 m altitude, and may be stored from −20 to +71 °C. It is designed to EN 61010-1, ANSI/ISA S82.01 and CAN/CSA C22.2 No. 1010.1.
Please tell us which pressure ranges you need to calibrate and whether test pumps should be included, so that we can offer the calibrator with the right modules.
726, 750PA6, 750PA27, 750PA7, 750PA8, 750PA9, 750PV3, 750PV4, 750PD2, 750PD3, 750PD10, 750PD4, 750PD5, 750PD50, 750PD6, 750PD7, 750PD27, 750R04, 750R06, 750R27, 750R07, 750R08, 750R29, 750R30, 750R31, 750RD5, 750RD6, 750RD27, 750P03, 750P02
Identify the instrument under test and the points to reproduce. State the required uncertainty, reference conditions and certificate format.
For an installed instrument, copy the complete ordering code and its suffixes from the nameplate. Identify the part that will remain in service and the part that needs to be replaced. A photograph of the connection face or terminal markings helps resolve differences between versions that share the same housing.
For a new installation, describe what the instrument must measure or control and where the reading will be used. Give the normal operating conditions separately from short peaks. Include units with every numerical value. A drawing with the available space, orientation and access for removal gives the supplier a practical starting point.
Add the quantity for each position and identify any accessories that belong with it. Where a particular drawing, material certificate or calibration record is required, mention it with the line item. Submit the list with the destination country so that the response can cover the requested configuration and delivery.