APLISENS AR250 / AR252 / AR253 / AR257 / AR258 / AR259
APLISENS

AR250–AR259 humidity and air quality transmitters

Humidity & air temperature

Manufacturer
APLISENS
Model
AR250 / AR252 / AR253 / AR257 / AR258 / AR259
Instrument type
Humidity & air temperature
Models
AR250, AR252, AR253, AR257, AR258, AR259
Humidity
0–100 % RH; ±2.5 % RH (0–90 %), ±3.5 % RH (90–100 %)
Temperature
−30 to +80 °C, ±0.5 °C
Pressure
300–1100 hPa, ±2 hPa
CO2
0–10,000 ppm; ±(40 ppm + 5 %) typical in 400–5000 ppm
Dust
1–1000 µg/m³; ±10 µg/m³ to 100 µg/m³, ±10 % above
VOC
0–9999 ppm; typically ±15 % of value
Response time
8–10 s humidity and temperature; 1 s other measurements
Calculated values
dew/frost point, absolute humidity
Analogue outputs
2 × 0/4–20 mA or 2 × 0/2–10 V
Digital interface
RS-485, Modbus RTU slave
Power supply
12–36 V DC
Housing
ABS / polycarbonate, IP65
Probe versions
integrated, external on 1.5 m cable, stainless steel pipe 140 or 240 mm
Request a quote
Review enquiry

The AR250, AR252, AR253, AR257, AR258 and AR259 are a family of digital transmitters for air quality monitoring. Depending on the model they measure humidity, temperature, atmospheric pressure, CO2, dust and volatile organic compounds (VOC), and pass the results on as analogue signals or over a digital interface.

The models differ in their measured parameters. The AR250 measures humidity; the AR252 humidity and temperature; the AR253 pressure, humidity and temperature; the AR257 CO2, humidity and temperature; the AR258 dust, pressure, humidity and temperature; and the AR259 VOC, humidity and temperature. The devices use integrated sensors and can also calculate derived values such as dew or frost point and absolute humidity, which can be assigned to the outputs.

Humidity is measured from 0 to 100 % RH with an accuracy of ±2.5 % RH between 0 and 90 % RH and ±3.5 % RH above 90 % RH. Temperature is measured from −30 to +80 °C with ±0.5 °C, and pressure from 300 to 1100 hPa with ±2 hPa. CO2 is measured from 0 to 10,000 ppm, typically ±(40 ppm + 5 %) in the 400 to 5000 ppm range. Dust is measured from 1 to 1000 µg/m³ with ±10 µg/m³ up to 100 µg/m³ and ±10 % above. VOC is measured from 0 to 9999 ppm, typically ±15 % of the value. Response time is 8 to 10 s for humidity and temperature and 1 s for the other measurements, at an airflow above 3.6 km/h.

Outputs are two 0/4–20 mA current outputs or two 0/2–10 V voltage outputs, plus an RS-485 interface with Modbus RTU (slave). Supply voltage is 12 to 36 V DC.

The housing is made of ABS and polycarbonate and rated IP65. An LCD display with keyboard is optional. Probe versions include an integrated probe, an external probe on a 1.5 m cable and a probe on a stainless steel pipe of 140 or 240 mm, with radial or back installation. Configuration is carried out through the PRG port with the AR955/AR956 programmer or with configuration software.

Applications named for the series are HVAC installations, offices and residential environments, warehousing, production, the food industry, pharmacy, medicine, gardening and laboratories. Please state the required parameters, the output type, the probe version and whether a display is needed when you enquire.

Part numbers · 6

AR250, AR252, AR253, AR257, AR258, AR259

Industry applications

Selection notes

Include the sensing target, mounting clearance and direction of travel or force. Record the connector pinout from the existing device.

For an installed instrument, copy the complete ordering code and its suffixes from the nameplate. Identify the part that will remain in service and the part that needs to be replaced. A photograph of the connection face or terminal markings helps resolve differences between versions that share the same housing.

For a new installation, describe what the instrument must measure or control and where the reading will be used. Give the normal operating conditions separately from short peaks. Include units with every numerical value. A drawing with the available space, orientation and access for removal gives the supplier a practical starting point.

Add the quantity for each position and identify any accessories that belong with it. Where a particular drawing, material certificate or calibration record is required, mention it with the line item. Submit the list with the destination country so that the response can cover the requested configuration and delivery.

Related instruments

Find equipment

DEVICEEU