HORIBA ASGU-370TS SPLIT
HORIBA

ASGU-370TS SPLIT portable calibration unit for gas analysers with zero gas generator

Process gas analysers & chromatographs

Manufacturer
HORIBA
Model
ASGU-370TS SPLIT
Instrument type
Process gas analysers & chromatographs
Type
portable one-component multipoint calibration system with zero gas generator
Gas generation
dilution (built-in cylinder), permeation, gas phase titration (GPT), UV radiation
Flow control
factory-calibrated digital thermal mass flow controllers
Span gas repeatability / linearity
±1 % / ±2 %
Temperature
50 °C, accuracy ±0.1 °C
Span points
up to 20 freely definable
Modules
up to 6; serial (ASGU-370S) or parallel (ASGU-370P)
Operation
touch-screen display; remote software
Interfaces
RS-232, Ethernet (Bayern-Hessen protocol)
Start-up / warm-up time
20 s / 30 min
Battery pack (UPS)
approx. 30 min
Power supply
230 V, 50 Hz
Case 1
600 × 400 × 155 mm (W × H × D), 12–16 kg
Case 2
600 × 400 × 115 mm (W × H × D), 8 kg
Standards
EN 14211, EN 14212, EN 14626
Request a quote
Review enquiry

The ASGU-370TS SPLIT is a portable, one-component multipoint calibration system for quality control of gas analysers in the laboratory and in the field. It belongs to the ASGU-370 series and includes a generator of zero gas. Its split design divides the system into two stackable cases for ergonomic transport.

The system produces zero gas and span gas dynamically. Depending on the installed modules, span gas is generated by dilution of gas from a built-in cylinder, by permeation, by gas phase titration (GPT) or by UV radiation. Factory-calibrated digital thermal mass flow controllers set the gas flows precisely; for certain technologies heated ovens or specific sources such as Hg lamps or permeation tubes are used. Up to 20 span points can be freely defined.

Span gas repeatability is ±1 % and span gas linearity ±2 %. The stated temperature is 50 °C with an accuracy of ±0.1 °C. The system starts up in 20 seconds and needs 30 minutes to warm up. Operating conditions are stored continuously.

Up to six modules can be combined. In the ASGU-370S they operate in series, and in the ASGU-370P up to six modules work in parallel as independent units. Rack-mountable variants in 19-inch 3U format are available.

Operation is via a touch-screen display, and remote software support is provided. RS-232 and Ethernet interfaces use the Bayern-Hessen protocol, which allows the calibrator to be controlled from a monitoring station. An internal data logger is optional. A battery pack acts as an uninterruptible power supply for approximately 30 minutes; mains supply is 230 V, 50 Hz.

Case 1 measures 600 × 400 × 155 mm (W × H × D) and weighs 12 to 16 kg; case 2 measures 600 × 400 × 115 mm and weighs 8 kg. The system supports calibration work related to EN 14211, EN 14212 and EN 14626, the standards for ambient air measurement of nitrogen oxides, sulphur dioxide and carbon monoxide.

Applications are quality control in the laboratory and in the field, calibration of gas analysers and industrial process gas analysis.

Please state the gas component to be generated, the required span points and the module technology, and whether the data logger or battery module is needed.

Industry applications

Selection notes

Identify the target gas, background gas and concentration units. State whether the sampling system is already installed or belongs in the enquiry.

For an installed instrument, copy the complete ordering code and its suffixes from the nameplate. Identify the part that will remain in service and the part that needs to be replaced. A photograph of the connection face or terminal markings helps resolve differences between versions that share the same housing.

For a new installation, describe what the instrument must measure or control and where the reading will be used. Give the normal operating conditions separately from short peaks. Include units with every numerical value. A drawing with the available space, orientation and access for removal gives the supplier a practical starting point.

Add the quantity for each position and identify any accessories that belong with it. Where a particular drawing, material certificate or calibration record is required, mention it with the line item. Submit the list with the destination country so that the response can cover the requested configuration and delivery.