EDT 101
Temperature

RTD sensors
The Balco Strap-On is a non-intrusive temperature sensor for measuring the surface temperature of pipes. It clamps directly onto the outside of the pipe, so it can be retrofitted to existing piping without breaking into the line. Typical uses are hydronic heating supply and return lines, chilled-water piping and chillers, cold storage facilities and high dew point or condensing environments; outside air temperature sensing is also listed among the applications.
The sensing element is a Balco resistance temperature detector with a PTC characteristic and a nominal resistance of 1000 Ω at 70 °F (21.1 °C); the temperature coefficient is 4618 ppm/°C. The element is encapsulated on the back of a 1.5-inch square copper plate, which maximises the thermal conductivity between the pipe and the sensor. Accuracy is ±1 % at 70 °F, the nominal response time is 8 seconds, and stability is ±0.05 % after 1000 hours at 150 °C. The recommended operating current is 5 mA.
The sensor is connected through two 22 AWG leads with etched PTFE insulation, colour-coded and not sensitive to polarity. It operates from −40 to 70 °C and at 10 to 95 % RH non-condensing, and can be stored from −40 to 85 °C.
Three enclosure versions are available for the termination: galvanised steel (A/BALCO-S-GD), ABS plastic (A/BALCO-S-PB) and a NEMA 4X weatherproof plastic enclosure (A/BALCO-S-4X). The sensor is fixed with adjustable hose clamps, offered for 2 to 5 in and 2 to 12 in pipe diameters; the series is listed for pipe sizes from 1.25 in. Heat transfer paste and a three-point NIST certificate are available as options. The sensors comply with CE, UKCA, RoHS2 and WEEE.
For best results, the pipe surface should be cleaned and thermal grease applied before mounting, and the sensor should then be insulated from the effects of ambient air.
Please send the pipe diameter, required enclosure and whether a calibration certificate is needed with your enquiry.
A/BALCO-S-GD, A/BALCO-S-PB, A/BALCO-S-4X, 114910, 124210, 137376, A/HOSE CLAMP-2-5", A/HOSE CLAMP-2-12", 142630, 142631, 102595, 140574
Match the sensing element, insertion length and wiring to the existing input. Separate the process temperature from the ambient temperature at the connection head.
For an installed instrument, copy the complete ordering code and its suffixes from the nameplate. Identify the part that will remain in service and the part that needs to be replaced. A photograph of the connection face or terminal markings helps resolve differences between versions that share the same housing.
For a new installation, describe what the instrument must measure or control and where the reading will be used. Give the normal operating conditions separately from short peaks. Include units with every numerical value. A drawing with the available space, orientation and access for removal gives the supplier a practical starting point.
Add the quantity for each position and identify any accessories that belong with it. Where a particular drawing, material certificate or calibration record is required, mention it with the line item. Submit the list with the destination country so that the response can cover the requested configuration and delivery.