Calibration

Temperature calibrators
BB705 is a very high temperature, laboratory-grade blackbody calibration source for testing and calibrating non-contact infrared temperature instruments. It provides a cavity at a precisely controlled temperature whose radiation the infrared thermometer under test views, and it serves as a reference standard for infrared pyrometers.
The blackbody cavity operates from 100 to 1046 °C (212 to 1915 °F). With a cavity emissivity of 0.99, the radiation emitted is very close to that of an ideal black body, so the infrared instrument can be checked without uncertainty from an unknown surface emissivity. The temperature of the cavity is measured by internal platinum thermocouple sensing elements and controlled by a built-in PID controller with autotune function and digital readout.
Stated accuracy is ±0.25 % of reading; ±1 °C, and stability is 0.4 °C (0.7 °F) per 8-hour period. A built-in NIST-traceable reference temperature display shows the actual cavity temperature. The large aperture of 44.5 mm (1.75 in) allows instruments with larger measuring spots to be calibrated, since the measuring spot of the instrument must be filled by the cavity opening.
The calibrator is fitted with an RS232 interface as standard for connection to a computer. The display is an LED type. The unit is of portable design and measures 600 mm in length and 520 mm in height. It operates at ambient temperatures from 0 to 35 °C.
Two versions are available according to the mains supply: BB705 for 110 V AC and BB705-230VAC for 230 V AC, both for 50/60 Hz.
Typical uses stated for the instrument are the calibration of non-contact infrared temperature instruments and use as a reference standard for infrared pyrometers in the laboratory. Please send the supply voltage and details of the infrared thermometers to be calibrated, including their temperature range and spot size, with your enquiry.
BB705, BB705-230VAC
Identify the instrument under test and the points to reproduce. State the required uncertainty, reference conditions and certificate format.
For an installed instrument, copy the complete ordering code and its suffixes from the nameplate. Identify the part that will remain in service and the part that needs to be replaced. A photograph of the connection face or terminal markings helps resolve differences between versions that share the same housing.
For a new installation, describe what the instrument must measure or control and where the reading will be used. Give the normal operating conditions separately from short peaks. Include units with every numerical value. A drawing with the available space, orientation and access for removal gives the supplier a practical starting point.
Add the quantity for each position and identify any accessories that belong with it. Where a particular drawing, material certificate or calibration record is required, mention it with the line item. Submit the list with the destination country so that the response can cover the requested configuration and delivery.