LDH313
Gas analysis

Flue gas & emissions analysers
CEM500 is an extractive online gas analyser for continuous emission monitoring (CEM). It measures several gas components in industrial exhaust streams simultaneously using UV spectroscopy, and it is also used for process and environmental measurements, water pollution monitoring, DeNOx applications in coal power plants and motor bench applications.
The analyser uses a xenon flash lamp as the light source and a CCD detector with 2048 pixels to record absorbance spectra with a resolution of 0.1 nm over a standard optical path of 240 mm. Fast Fourier Transform (FFT) algorithms give high selectivity for gases with periodic absorption structures such as NH3, SO2 and NO. The internal gas circuit is heated to 190 °C, so hot and humid combustion gases can be analysed directly. Measurement response is between 200 ms and 5 s.
The sample gas can be supplied at 0 to 2 bar absolute, with a flow of 0.1 to 10 l/min and temperatures from ambient up to 400 °C; zero gas is supplied at the same pressure and flow ranges. The lamp lifetime is three years, which keeps maintenance low.
Results are shown on a colour TFT LCD screen with 640 × 480 pixels, and up to 5000 measurements are stored. Standard communication includes RS-232 and USB with Modbus. Optional modules extend the analyser: an isolated 4–20 mA input module (IN4-20), a double logic input module with isolated 0–24 V DC inputs (LOGIC500), an isolated active 4–20 mA output module with a maximum load of 500 Ω (OUT4-20-500), a relay output module with 2 A/220 V contacts (RELAY500), an internal Wi-Fi interface to IEEE 802.11b (WIFI500) and an internal 10Base-T Ethernet interface to IEEE 802.3 (ETHER500). A membrane sampling pump with heated head, about 6 l/min and heated to 190 °C, is available as PUMP500. Power supply is 110–240 V AC, 50–60 Hz, 1000 VA.
The analyser is housed in a wall-mounted IP65 stainless steel enclosure measuring 345 × 525 × 260 mm and weighs 30 kg. Gas connections are stainless steel 316 compression fittings for ¼ in (6.4 mm) OD tube. Operating limits are 0 to 40 °C and less than 90 % relative humidity. The product conforms to EN/IEC 61010-1 and EN/IEC 61326 and complies with RoHS, REACH and EMC requirements.
Please send the gas components and ranges to be measured, the sample gas conditions and the required outputs and options with your enquiry.
PUMP500, IN4-20, LOGIC500, OUT4-20-500, RELAY500, WIFI500, ETHER500
Identify the target gas, background gas and concentration units. State whether the sampling system is already installed or belongs in the enquiry.
For an installed instrument, copy the complete ordering code and its suffixes from the nameplate. Identify the part that will remain in service and the part that needs to be replaced. A photograph of the connection face or terminal markings helps resolve differences between versions that share the same housing.
For a new installation, describe what the instrument must measure or control and where the reading will be used. Give the normal operating conditions separately from short peaks. Include units with every numerical value. A drawing with the available space, orientation and access for removal gives the supplier a practical starting point.
Add the quantity for each position and identify any accessories that belong with it. Where a particular drawing, material certificate or calibration record is required, mention it with the line item. Submit the list with the destination country so that the response can cover the requested configuration and delivery.