LDH313
Gas analysis

Flue gas & emissions analysers
CEM500-L is an extractive online gas analyser for continuous emission monitoring systems (CEMS), designed to meet strict low-level emission limits. It belongs to the CEM500 family and measures several gas components simultaneously by UV spectroscopy.
The analyser passes the sample through a 1200 mm multi-path flow cell. Light from a xenon flash lamp is analysed by a high-resolution spectrograph reading a 2048-pixel CCD. Each gas is identified by its specific absorption spectrum, and FFT algorithms provide high selectivity and sensitivity without interference from CO, CO2 or CH4. Zero compensation is automatic, and the heated gas circuit can be set from 40 to 240 °C. The lamp lasts approximately three years in continuous operation.
Ranges are 0–10 ppm NH3 (0–7 mg/m³, detection limit 0.1), 0–80 ppm NO (0–100 mg/m³, detection limit 0.5), 0–30 ppm SO2 (0–100 mg/m³), 0–200 ppm H2S and NO2 (0–400 mg/m³ each), 0–20 ppm CS2 and benzene (0–60 mg/m³), 0–5 ppm toluene and xylene (0–20 mg/m³) and 0–20 ppm NCl3 (0–100 mg/m³). Sample and zero gas are supplied at 0–2 bar absolute and 0.1 to 10 l/min; sample temperature may be up to 400 °C.
Results are shown on a colour TFT LCD screen with LED backlight (640 × 480 pixels), and up to 5000 measurements are stored and can be exported via USB. Options include one to eight opto-isolated 4–20 mA outputs, one to four programmable relay contacts, 4–20 mA input and logic input modules, a Wi-Fi interface (WIFI500, IEEE 802.11b), an Ethernet interface (ETHER500, IEEE 802.3), a phone modem (MTI133), a GSM modem and an internal membrane pump with heated head (PUMP500). Power supply is 110–240 V AC, 50–60 Hz, 1000 VA.
The enclosure is coated stainless steel to IP65, measuring 345 × 525 × 260 mm, and the analyser weighs 25 kg. Gas connections are stainless steel 316 compression fittings for ¼ in (6.4 mm) OD tube. Operating limits are 0 to 40 °C and below 90 % relative humidity. The analyser conforms to EN/IEC 61010-1 and EN/IEC 61326 and to RoHS, REACH, EMC and R&TTE requirements.
Applications include CEMS, chemical processes requiring low H2S values, pharmaceutical and biopharmaceutical production, semiconductor manufacturing, environmental impact assessment, carbon capture, hydrogen energy, hydrogen and ammonia engine evaluation and fuel cell evaluation.
Please specify the gases and ranges, the sample conditions and the outputs and communication options required.
PUMP500, IN4-20-500, LOGIC500, OUT4-20-500, RELAY500, WIFI500, ETHER500, MTI133, GSM
Identify the target gas, background gas and concentration units. State whether the sampling system is already installed or belongs in the enquiry.
For an installed instrument, copy the complete ordering code and its suffixes from the nameplate. Identify the part that will remain in service and the part that needs to be replaced. A photograph of the connection face or terminal markings helps resolve differences between versions that share the same housing.
For a new installation, describe what the instrument must measure or control and where the reading will be used. Give the normal operating conditions separately from short peaks. Include units with every numerical value. A drawing with the available space, orientation and access for removal gives the supplier a practical starting point.
Add the quantity for each position and identify any accessories that belong with it. Where a particular drawing, material certificate or calibration record is required, mention it with the line item. Submit the list with the destination country so that the response can cover the requested configuration and delivery.