DwyerOmega CL-355A
DwyerOmega

CL-355A portable dry block temperature calibrator

Temperature calibrators

Manufacturer
DwyerOmega
Model
CL-355A
Instrument type
Temperature calibrators
Temperature range
5 °C above ambient to 400 °C (752 °F)
Accuracy
±0.6 °C
Stability after 15 min
±0.15 °C
Well-to-well uniformity
0.015 to 0.025 °C
Display resolution
0.1°
Heat-up time (20 to 400 °C)
12 min
Cool-down time (400 to 100 °C)
21 min
Insert
machined aluminium; multi-well 1/8, 3/16, 1/4, 5/16, 3/8″ included; 10 formats available
Insert immersion depth
114 mm
Power
900 W; 120 V AC or 240 V AC, 50/60 Hz
Dimensions
222 × 203 × 203 mm
Weight
4 kg
Approvals
CE, RoHS
Request a quote
Review enquiry

The CL-355A is a compact, portable dry block temperature calibrator. It provides a constant-temperature source for checking and calibrating temperature sensors both for standard laboratory work and for calibration in the field.

Heat is transferred to the sensor under test through a machined aluminium insert block placed in the heated well. The sensors are inserted into bores of the insert and brought to the set temperature. The calibrator covers a range from 5 °C (9 °F) above ambient up to 400 °C (752 °F).

The stated accuracy is ±0.6 °C (1.1 °F), and stability after 15 minutes is ±0.15 °C (0.27 °F). Well-to-well uniformity is 0.015 to 0.025 °C, so several sensors in different bores see practically the same temperature. The display resolution is 0.1°. The insert immersion depth is 114 mm.

Thermal cycling is fast: heating from 20 to 400 °C takes 12 minutes, and cooling from 400 to 100 °C takes 21 minutes with automatic fan cooling. An independent over-temperature cutout provides safety protection. The heater rating is 900 W, and the calibrator is available for 120 V AC (CL-355A) or 240 V AC (CL-355A-230), 50/60 Hz.

A multi-well insert with bores for 1/8, 3/16, 1/4, 5/16 and 3/8 in sensors is included, and ten insert formats are available in total, so the block can be matched to the sensor diameters being calibrated. A carrying case (CL-350A-CASE) is available for transport. The unit measures 222 × 203 × 203 mm and weighs 4 kg (9 lb). It is CE marked and RoHS compliant, and a NIST-traceable calibration certificate can be supplied.

Please send the diameters and number of sensors to be calibrated, the temperature range and the mains voltage with your enquiry, and state whether a calibration certificate is needed.

Part numbers · 12

CL-355A, CL-355A-230, CL-350A-CASE, CL701B, CL703B, CL704B, CL705B, CL706B, CL707B, CL708B, CL709B, CL710B

Industry applications

Selection notes

Identify the instrument under test and the points to reproduce. State the required uncertainty, reference conditions and certificate format.

For an installed instrument, copy the complete ordering code and its suffixes from the nameplate. Identify the part that will remain in service and the part that needs to be replaced. A photograph of the connection face or terminal markings helps resolve differences between versions that share the same housing.

For a new installation, describe what the instrument must measure or control and where the reading will be used. Give the normal operating conditions separately from short peaks. Include units with every numerical value. A drawing with the available space, orientation and access for removal gives the supplier a practical starting point.

Add the quantity for each position and identify any accessories that belong with it. Where a particular drawing, material certificate or calibration record is required, mention it with the line item. Submit the list with the destination country so that the response can cover the requested configuration and delivery.