
LDH313
Gas analysis

Process gas analysers & chromatographs
The FRP-4 / CPA-4 is an online physical property analyser that measures the freezing point and the cloud point of a product directly in the process stream. It is intended for products such as jet fuels for the aviation industry, gasoils and diesel products, and is built as an explosion-protected assembly for hazardous areas.
Measurement is based on optical turbidity detection while the sample is cooled. The cloud point is determined from the reduction of light intensity as crystals begin to form; the freezing point is identified when the crystals dissolve. The method complies with ASTM D2386, ASTM D1015, DIN ISO 3013, ASTM D7153-05 and ASTM D7154-05. For cloud point applications, typically no chiller is required; an optional chiller FKS-KWS is available. Integrated failure diagnosis and self-monitoring are part of the system.
The sample must arrive filtered to 50 µm and free of suspended water, with a viscosity of up to 37 cSt at inlet temperature, a flow of about 5 to 30 l/h, a pressure of 2 to 3 bar and a temperature of 5 to 15 °C and at least 15 K above the expected cloud point. It leaves at atmospheric pressure. Up to two sample streams and one validation stream can be connected via tube fittings, typically 6, 12 or 18 mm.
Up to eight 4–20 mA outputs (1000 Ω) transmit freezing point, cloud point and other values on request, with active isolated outputs available. Digital signals cover alarm and ready status, and inputs for stream selection, validation request and reset are available. Communication includes Modbus RTU via RS-485, RS-422 or fibre optic and Modbus TCP, with Ethernet over VDSL or fibre optic. Operation is via a 19-inch touch display running Windows 10 Enterprise LTSB with cyber security features; remote access is available on request.
Without the chiller the analyser is marked ATEX Ex II 2G Ex h IIC T4 Gb and IECEx Ex h IIC T4 Gb; with the chiller FKS-KWS the marking is Ex II 2G Ex h IIB+H2 T4 Gb. It operates at ambient temperatures from 5 to 35 °C, has IP54 protection (comparable with NEMA 13), measures approximately 1140 × 2050 × 710 mm and weighs about 380 kg without options.
Options include remote access, a validation report, freely programmable digital and analogue inputs, a sample conditioning system, a sample recovery system, a fast loop pump and sample probes. Consulting, engineering, after-sales, returns and repair services are available.
Please send the product type, the expected freeze or cloud point, the sample conditions and the area classification with your enquiry.
Identify the target gas, background gas and concentration units. State whether the sampling system is already installed or belongs in the enquiry.
For an installed instrument, copy the complete ordering code and its suffixes from the nameplate. Identify the part that will remain in service and the part that needs to be replaced. A photograph of the connection face or terminal markings helps resolve differences between versions that share the same housing.
For a new installation, describe what the instrument must measure or control and where the reading will be used. Give the normal operating conditions separately from short peaks. Include units with every numerical value. A drawing with the available space, orientation and access for removal gives the supplier a practical starting point.
Add the quantity for each position and identify any accessories that belong with it. Where a particular drawing, material certificate or calibration record is required, mention it with the line item. Submit the list with the destination country so that the response can cover the requested configuration and delivery.