Surent G480
Flow

Flow switches & indicators
FS10A is a universal flow switch and monitor developed specifically for gas and liquid process analyser sampling systems. It confirms that sample is flowing to the analyser and signals when flow falls below or rises above a set point. A single configuration covers all flow ranges with a turndown of up to 100:1, and it offers particularly good sensing of low flows.
The instrument uses thermal dispersion flow measurement with equal mass sensing. Heat transfer from a heated sensor to the passing medium changes with flow rate, and this change is evaluated by the electronics. The sensor has no moving parts, so there is nothing to wear and no maintenance. An LED array shows the flow rate and the switch trip status, and set points are adjusted with push-buttons or over the RS232C interface.
For gases the flow range extends from 10 to 100,000 cc/min (0.02 to 200 SCFH), for liquids from 0.70 to 750 cc/min (0.01 to 12 GPH), with a turndown of 5:1 to 100:1. Repeatability is ±0.5 % of reading. Standard temperature compensation covers ±16 °C, extended compensation ±55 °C. The temperature coefficient is at most ±0.05 % of reading per °C up to 260 °C for gas and ±0.367 % of reading per °C up to 121 °C for liquids. Standard operating temperature is −40 to 121 °C.
Output choices include an N-channel open collector rated 100 mA, an SPDT relay rated 1 A at 24 V DC or 120 V AC (DC only for ATEX), a 4–20 mA trend output and RS232C serial I/O. Wetted parts are 316L stainless steel with Hastelloy C-22 thermowells, or optionally an all-Hastelloy C-22 probe assembly. The process connection is 1/4" NPT, compatible with 3/8" and 1/2" tube tees and with an SP76 adapter for NeSSI modular sampling systems.
The switch is approved as nonincendive for Class I Division 2 and marked II 3 G Ex ec IIC T4 Gc and II 3 D Ex tc IIIC T81 °C Dc, with ATEX, IECEx, FM and CE approvals, and it is SIL 2 rated to IEC 61508. Enclosure ratings from IP64 to IP67 and NEMA 4X apply depending on version. Integral or remote configurations are available, as are sanitary, nuclear and high-temperature options.
Please send the medium, expected flow range, process connection, required outputs and the area classification of the sampling system.
019897-01
State the minimum, normal and peak flow with units. Pipe bore, fluid properties and the available straight run affect the choice of measuring principle.
For an installed instrument, copy the complete ordering code and its suffixes from the nameplate. Identify the part that will remain in service and the part that needs to be replaced. A photograph of the connection face or terminal markings helps resolve differences between versions that share the same housing.
For a new installation, describe what the instrument must measure or control and where the reading will be used. Give the normal operating conditions separately from short peaks. Include units with every numerical value. A drawing with the available space, orientation and access for removal gives the supplier a practical starting point.
Add the quantity for each position and identify any accessories that belong with it. Where a particular drawing, material certificate or calibration record is required, mention it with the line item. Submit the list with the destination country so that the response can cover the requested configuration and delivery.