HORIBA MEXA-1170HFID
HORIBA

MEXA-1170HFID heated FID total hydrocarbon analyser

Flue gas & emissions analysers

Manufacturer
HORIBA
Model
MEXA-1170HFID
Instrument type
Flue gas & emissions analysers
Measuring principle
Heated flame ionisation detection (HFID)
Component
Total hydrocarbons (THC)
Ranges
0–10 ppmC to 0–200,000 ppmC, max. 14 ranges
Response time (T90)
Within 1.5 s
Repeatability
Within ±1.0 % FS
Linearity
Within ±1.0 % FS
Drift
Within ±2.0 % FS per 8 h
Noise
Within ±1.0 % FS
Sample line temperature
113 ±8 °C or 191 ±10 °C (specify on order)
Sample gas
0–30 kPa; approx. 2 L/min
Warm-up time
Within 2 h after ignition
Outputs
RS-232C; analogue 10 V optional
Ambient conditions
5 to 40 °C, < 80 % RH
Power
100/115/120/230 V AC ±10 %, 50/60 Hz, approx. 3 kVA
Dimensions / weight
464 × 652 × 185 mm; approx. 25 kg
Request a quote
Review enquiry

The MEXA-1170HFID is a heated-type total hydrocarbon (THC) analyser based on flame ionisation detection. It is designed for engine testing of automobiles and components and analyses direct (raw) exhaust, CVS/tunnel-diluted exhaust and evaporative emission samples from SHED tests. It can be used with gasoline, diesel, methanol, LPG and CNG engines.

In a flame ionisation detector, hydrocarbons in the sample are burned in a hydrogen flame, producing ions whose current is proportional to the carbon content; the result is expressed in ppmC. In the heated version, the sample path is kept at a controlled elevated temperature. The sample line temperature is 113 ±8 °C or 191 ±10 °C and must be specified when ordering. The sampler, the HFID detector and the control unit are combined in one slim, compact case for stand-alone operation.

Up to 14 measuring ranges can be selected, from 0–10 ppmC to 0–200,000 ppmC. Response time (T90) is within 1.5 seconds, measured by switching between zero and span lines. Repeatability, linearity and noise are each within ±1.0 % of full scale, and drift is within ±2.0 % of full scale per 8 hours at 20 ±5 °C ambient. The sample gas pressure must be between 0 and 30 kPa, and the sample flow is approximately 2 L/min. Warm-up takes up to 2 hours after ignition of the detector.

Measured data are transferred to a remote computer via the RS-232C digital output; an analogue 10 V output module is available as an option. Digital inputs and outputs are provided.

The analyser operates on 100, 115, 120 or 230 V AC ±10 %, 50/60 Hz single phase, with a consumption of approximately 3 kVA. It measures 464 × 652 × 185 mm (W × D × H) and weighs about 25 kg. The ambient range is 5 to 40 °C at less than 80 % RH. The instrument conforms to RoHS, REACH and EMC requirements.

Accessories include the heated filter HF-04 for diesel and direct-injection gasoline engines, the heated filter HF-05 for two-cycle and general-purpose engines, heated lines of 2, 3, 4 or 6 metres, a temperature controller and the analogue output module.

When requesting a quotation, please state the sample line temperature, the engine types to be tested, the required heated filter and line length, and the supply voltage.

Part numbers · 1

MEXA-1170HFID

Industry applications

Selection notes

Identify the target gas, background gas and concentration units. State whether the sampling system is already installed or belongs in the enquiry.

For an installed instrument, copy the complete ordering code and its suffixes from the nameplate. Identify the part that will remain in service and the part that needs to be replaced. A photograph of the connection face or terminal markings helps resolve differences between versions that share the same housing.

For a new installation, describe what the instrument must measure or control and where the reading will be used. Give the normal operating conditions separately from short peaks. Include units with every numerical value. A drawing with the available space, orientation and access for removal gives the supplier a practical starting point.

Add the quantity for each position and identify any accessories that belong with it. Where a particular drawing, material certificate or calibration record is required, mention it with the line item. Submit the list with the destination country so that the response can cover the requested configuration and delivery.