DwyerOmega O2S-FR-T2
DwyerOmega

O2S zirconia oxygen sensor, probe version

Oxygen analysers

Manufacturer
DwyerOmega
Model
O2S-FR-T2
Instrument type
Oxygen analysers
Sensing element
zirconium dioxide (ZrO₂), integral heater
Measured value
oxygen partial pressure
Range
2 mbar to 3 bar ppO₂
Accuracy
5 mbar max.
Response time (10–90 %)
< 15 s standard; < 4 s fast
Internal operating temperature
700 °C
Gas temperature
−100 to +250 °C standard; to +400 °C high-temperature
Gas flow rate
0–10 m/s
Max. pressure
3 bar
Mounting thread
M18 × 1.5 or 3/4 in NPT
Housing
steel, long probe housing
Electronics
external interface board (analogue or digital outputs)
Request a quote
Review enquiry

This zirconia oxygen sensor measures the partial pressure of oxygen in gas mixtures with a zirconium dioxide sensing element. It does not need a reference gas and does not require temperature stabilisation of the sample, which makes installation simple. The technology is non-depleting, so the sensor has a long service life.

An integral heater keeps the sensing element at its internal operating temperature of 700 °C. The sensor measures oxygen partial pressure from 2 mbar up to 3 bar with an accuracy of 5 mbar. Response time from 10 to 90 % is below 15 seconds for the standard version and below 4 seconds for the fast-response version. Permissible gas temperature is from −100 to +250 °C for the standard design and up to +400 °C for the high-temperature version, with gas flow velocities up to 10 m/s. The sensor withstands repetitive accelerations of 5 g and incidental accelerations of 30 g.

The probe version has a long steel housing for insertion into ducts, pipes and process chambers. It is screwed in with an M18 × 1.5 metric thread or a 3/4 in NPT thread; a compact short-housing version is also available. Lead wires are PTFE-insulated, with a connector or with cable lengths of 0.15, 0.3 or 1.1 m.

For operation the sensor is connected to an external interface board that controls the heater and converts the sensor signal into an output such as 4–20 mA, 0–10 V or a digital signal via RS-232, RS-485 or Modbus. The sensor is CE and UKCA marked.

Typical applications are laboratory measurements, combustion control and boiler and furnace efficiency optimisation, automotive emissions testing, oxygen generation, fuel tank inerting systems in aircraft, agricultural processes and composting, bakery ovens, biomass monitoring and low-sulphur coal analysis. Please state the gas temperature, required response time, mounting thread and the interface board or output you need.

Part numbers · 4

O2S-FR-T2-34NPTA, O2S-FR-T2-18A, O2S-FR-T2-18B, O2S-FR-T2-18C

Industry applications

Selection notes

Identify the target gas, background gas and concentration units. State whether the sampling system is already installed or belongs in the enquiry.

For an installed instrument, copy the complete ordering code and its suffixes from the nameplate. Identify the part that will remain in service and the part that needs to be replaced. A photograph of the connection face or terminal markings helps resolve differences between versions that share the same housing.

For a new installation, describe what the instrument must measure or control and where the reading will be used. Give the normal operating conditions separately from short peaks. Include units with every numerical value. A drawing with the available space, orientation and access for removal gives the supplier a practical starting point.

Add the quantity for each position and identify any accessories that belong with it. Where a particular drawing, material certificate or calibration record is required, mention it with the line item. Submit the list with the destination country so that the response can cover the requested configuration and delivery.

Related instruments

Find equipment