HORIBA PG-300 series
HORIBA

PG-300 series portable multi-component flue gas analyser

Flue gas & emissions analysers

Manufacturer
HORIBA
Model
PG-300 series
Instrument type
Flue gas & emissions analysers
Measured components
NOx, SO2, CO, CO2, O2, CH4, N2O (depending on model)
Measuring principles
cross-modulation CLD (NOx), cross-modulation NDIR / NDIR (SO2, CO, CH4); O2 galvanic, zirconia or paramagnetic
NOx ranges
0–25 to 0–2500 ppm
SO2 ranges
0–200 / 500 / 1000 / 3000 ppm
CO ranges
0–200 / 500 / 1000 / 2000 / 5000 ppm
CO2 / O2 ranges
0–10 / 20 / 30 vol% / 0–5 / 10 / 25 vol%
CH4 / N2O ranges
0–2000 / 5000 ppm / 0–100 to 0–5000 ppm
Linearity
±2.0 % of full scale
Repeatability
±0.5 % FS (NOx ≥ 100 ppm, CO ≥ 1000 ppm ranges), otherwise ±1.0 % FS
Response time
≤ 45 s; SO2 ≤ 180 s
Warm-up time
30 min
Outputs and interfaces
DC 4–20 mA; 0–1 V optional; Ethernet; SD card slot
Power
AC 100–240 V ±10 %, 50/60 Hz; approx. 160 VA steady state, max. 220 VA
Dimensions
300 × 520 × 260 mm with side guards; 260 × 520 × 260 mm without
Weight
approx. 13–15 kg
Request a quote
Review enquiry

The PG-300 series is a compact portable gas analyser for field and laboratory use. Depending on the model, it measures up to five gas components at the same time, including the greenhouse gases methane (CH4) and nitrous oxide (N2O) alongside NOx, SO2, CO, CO2 and O2. The series comprises models PG-320, PG-324, PG-325, PG-330, PG-335, PG-337, PG-340, PG-344, PG-344 (Carbon Neutral) and PG-350.

Different detection methods are combined in one instrument. NOx is measured by cross-modulation chemiluminescence detection (CLD); SO2, CO and CH4 are measured by non-dispersive infrared absorption (NDIR), partly in cross-modulation form. For O2, galvanic, zirconia or paramagnetic detectors can be chosen. In the cross-modulation method, sample gas and reference gas are fed alternately, which reduces span drift.

Ranges are selectable per component: NOx from 0–25 up to 0–2500 ppm, SO2 0–200 to 0–3000 ppm, CO 0–200 to 0–5000 ppm, CO2 0–10, 20 or 30 vol%, O2 0–5, 10 or 25 vol%, CH4 0–2000 or 5000 ppm, and N2O 0–100 to 0–1000 ppm (C range) or 0–500 to 0–5000 ppm (D range). Linearity is ±2.0 % of full scale. Repeatability is ±0.5 % of full scale for NOx ranges from 100 ppm and CO ranges from 1000 ppm, and ±1.0 % otherwise. Drift is ±1.0 % of full scale per day (±2.0 % for SO2). Response time is 45 s or less, 180 s or less for SO2; the warm-up takes 30 minutes and the sample flow is approximately 0.5 l/min.

Measured values are available as DC 4–20 mA signals (non-insulated), with 0–1 V as an option, and data can be stored on an SD card or transferred via Ethernet. The analyser runs on AC 100–240 V ±10 %, 50/60 Hz, and draws approximately 160 VA in steady state (maximum 220 VA); power consumption has been reduced by 35 %.

The unit measures 300 × 520 × 260 mm with its shock-resistant side guards (260 mm wide without) and weighs approximately 13–15 kg. It is designed for 0–40 °C and up to 80 % RH. Options include an IP42 version with a special rear cover, an electronic cooler, a drain separator and a drain pot. Approvals listed are TÜV (EU), MCERTS (UK), GOST (Russia) and approvals for China, Japan and Korea.

Applications include stack emission testing in thermal, waste-to-energy and biomass power generation, iron and steel production, oil refineries and petrochemical plants, SO2 monitoring for flue gas desulphurisation, GHG emission monitoring and research, catalyst performance evaluation, fuel cells, process gas stream control and environmental impact assessment.

Please state the components and ranges you need, the preferred O2 detector and whether the IP42 version is required.

Part numbers · 10

PG-320, PG-325, PG-330, PG-335, PG-337, PG-340, PG-350, PG-324, PG-344, PG-344 (Carbon Neutral)

Industry applications

Selection notes

Identify the target gas, background gas and concentration units. State whether the sampling system is already installed or belongs in the enquiry.

For an installed instrument, copy the complete ordering code and its suffixes from the nameplate. Identify the part that will remain in service and the part that needs to be replaced. A photograph of the connection face or terminal markings helps resolve differences between versions that share the same housing.

For a new installation, describe what the instrument must measure or control and where the reading will be used. Give the normal operating conditions separately from short peaks. Include units with every numerical value. A drawing with the available space, orientation and access for removal gives the supplier a practical starting point.

Add the quantity for each position and identify any accessories that belong with it. Where a particular drawing, material certificate or calibration record is required, mention it with the line item. Submit the list with the destination country so that the response can cover the requested configuration and delivery.