HORIBA PoliSpectra RPR
HORIBA

PoliSpectra RPR rapid Raman plate reader for high-throughput screening

Optical & Raman spectrometers

Manufacturer
HORIBA
Model
PoliSpectra RPR
Instrument type
Optical & Raman spectrometers
Technique
Raman spectroscopy, non-destructive
Laser wavelength
532 nm or 785 nm
Laser power
up to 150 mW (532 nm); up to 500 mW (785 nm); software controlled
Resolution
1.7 cm⁻¹/pixel (532 nm); 1.6 cm⁻¹/pixel (785 nm)
Spectral coverage
400–3,150 cm⁻¹ (532 nm); 200–3,400 cm⁻¹ (785 nm)
Detector
deep-cooled scientific CCD or sCMOS, below −25 °C TE cooling from +20 °C ambient
Reading speed
96 wells in less than 1 minute (532 nm, 500 ms exposure per well)
Plate formats
96 or 384 wells; custom wells on request
Self-calibration
automatic, programmable timing
Door
motorised, for robot compatibility
Server control
OPC UA, REST API
Plate heater
included, for live reaction monitoring
Footprint
71.6 × 48.8 cm
Compliance
RoHS, REACH, EMC
Request a quote
Review enquiry

PoliSpectra RPR is a rapid Raman plate reader designed for high-throughput screening (HTS) in the pharmaceutical and biopharmaceutical sector. It records Raman spectra from the wells of a microplate one after another and is built to work inside automated laboratories, together with liquid handling systems and robots.

Raman spectroscopy illuminates the sample with a laser and analyses the small part of the scattered light that is shifted in wavelength by molecular vibrations. The resulting spectrum is a chemical fingerprint of the sample, obtained without contact and without destroying it. The reader uses a deep-cooled scientific camera, either CCD or sCMOS, cooled thermoelectrically to below −25 °C from a +20 °C ambient.

Two laser wavelengths are available. With 532 nm the laser power is up to 150 mW, the resolution 1.7 cm⁻¹ per pixel and the spectral coverage 400 to 3,150 cm⁻¹. With 785 nm the power is up to 500 mW, the resolution 1.6 cm⁻¹ per pixel and the coverage 200 to 3,400 cm⁻¹. Laser power is adjusted in software. A full 96-well plate is read in less than one minute at 532 nm with an exposure time of 500 ms per well. Well-to-well crosstalk is stated to be ultra-low, and well-to-well signal reproducibility is stated to be excellent. Self-calibration runs automatically at programmable intervals.

Plates with 96 or 384 wells are supported, and custom well formats are available on request. A plate heater is included, so reactions can be monitored live in the plate.

For automation, the reader has a motorised door that a robot arm can use for loading, and it is controlled as a server through OPC UA or a REST API. It is compatible with liquid handling systems. The footprint is 71.6 × 48.8 cm. The instrument complies with RoHS and REACH, and EMC conformity is documented.

Listed applications are vaccine development, bio-fermentation research and development, drug content quality control and high-throughput screening in the pharmaceutical industry.

Please let us know the required laser wavelength, detector type, plate format and the automation system to which the reader will be connected.

Industry applications

Selection notes

Describe the sample matrix and analytical method. Include the required measurement interval, sample interface and available bench space.

For an installed instrument, copy the complete ordering code and its suffixes from the nameplate. Identify the part that will remain in service and the part that needs to be replaced. A photograph of the connection face or terminal markings helps resolve differences between versions that share the same housing.

For a new installation, describe what the instrument must measure or control and where the reading will be used. Give the normal operating conditions separately from short peaks. Include units with every numerical value. A drawing with the available space, orientation and access for removal gives the supplier a practical starting point.

Add the quantity for each position and identify any accessories that belong with it. Where a particular drawing, material certificate or calibration record is required, mention it with the line item. Submit the list with the destination country so that the response can cover the requested configuration and delivery.