EDT 96
Pressure

Pressure transmitters
PX409-26BI is a high-accuracy pressure transducer for barometric pressure, measuring from 26 to 32 inches of mercury. It belongs to a family of precision transducers based on a silicon sensing element and produces a two-wire 4–20 mA output that can be connected directly to PLCs, data loggers, controllers and indicators.
Combined linearity, hysteresis and repeatability are within ±0.08 % of full scale (best straight line), which makes the unit suitable as a reference for test stands and for applications where small changes of atmospheric pressure must be detected. The narrow 6 inHg span concentrates the full output range on the pressures that actually occur at ground level, so the resolution of the measurement is high. The sensor has a bandwidth from DC to 1 kHz and a response time of less than one millisecond.
The transducer is supplied with 9 to 30 V DC. It has a 1/4 in NPT male process connection and a 2 m cable for the electrical connection. The housing is rated IP67, insulation resistance is 100 MΩ at 50 V DC, and the device operates from −45 to 115 °C. Compound ranges are calibrated in the positive direction only.
The same family is available with other options when a different version is needed: higher accuracy grades of ±0.05 % and ±0.03 %; electrical connections by cable with conduit, DIN 43650 C plug, MIL bayonet connector, hermetic pins or M12 4-pin connector; outputs of 0–5 V, 0–10 V, ±5 V, ±10 V, 10 mV/V, RS-485 or USB; pressure types gauge, compound, vacuum, absolute and sealed gauge, with ranges from 10 inH₂O to 3500 psi; NPT or BSPP process connections.
Typical uses are barometric compensation of other pressure and level measurements, weather and environmental stations, end-of-line testers, leak detection systems, test stands and process control. Please indicate whether you need exactly this barometric version or a different range, output or connection from the series.
PX409-26BI
Specify whether the reference is gauge, absolute or differential pressure. Include the normal span and the largest pressure the instrument may see.
For an installed instrument, copy the complete ordering code and its suffixes from the nameplate. Identify the part that will remain in service and the part that needs to be replaced. A photograph of the connection face or terminal markings helps resolve differences between versions that share the same housing.
For a new installation, describe what the instrument must measure or control and where the reading will be used. Give the normal operating conditions separately from short peaks. Include units with every numerical value. A drawing with the available space, orientation and access for removal gives the supplier a practical starting point.
Add the quantity for each position and identify any accessories that belong with it. Where a particular drawing, material certificate or calibration record is required, mention it with the line item. Submit the list with the destination country so that the response can cover the requested configuration and delivery.