Gas analysis

Oxygen analysers
The SERVOTOUGH Oxy 1900 is a digital paramagnetic oxygen analyser for use in hazardous areas classified as Zone 1 or Division 1. It measures the oxygen content of process gases in percent and is offered in configurations for general purpose, mid-level and full process oxygen measurement, including hydrogen-resilient options.
The analyser uses paramagnetic technology. Oxygen is strongly attracted by a magnetic field, unlike most other gases, and this property is used to determine its concentration. The measurement does not consume a sensing element, so it is stable and non-depleting. A heated sample compartment keeps the measuring cell at a stable temperature, and an internal pressure compensation option provides a 20:1 improvement in the effect of pressure changes for tighter process control.
The sample flow is 150 to 250 ml/min, nominally 200 ml/min. A bypass option with 50 to 70 l/h (60 l/h nominal) is available. The sample inlet is 1/4 in NPT female, 1/4 in OD compression or 6 mm OD compression. A dedicated flow sensor monitors the sample flow for enhanced safety. An auto-validation and calibration option reduces hands-on maintenance.
Supply voltage is 100 to 120 V AC or 220 to 240 V AC. Outputs include 4–20 mA as well as digital communication via Modbus RTU over RS485 and Modbus TCP over Ethernet.
The analyser carries ATEX, IECEx, UKEX and cCSAus certification for hazardous areas, as well as SIL, CE and RoHS. Standards referenced include BS EN 61010, IEC 61439-2 and IEEE C37.90, and for marine use IEC 60092-504, IACS UR E10, MARPOL Annex VI, the IGC Code and SOLAS. Versions are designated SV01, SV02, SV03, SV04 and UC. Support from design to installation of the analyser system is available.
Applications named are process control, feedstock purity, safety-critical oxidation, vapour recovery, flare stack analysis and marine vapour control. Please describe the sample composition, the oxygen range, the area classification and the required outputs with your enquiry.
SV01, SV02, SV03, SV04, UC
Identify the target gas, background gas and concentration units. State whether the sampling system is already installed or belongs in the enquiry.
For an installed instrument, copy the complete ordering code and its suffixes from the nameplate. Identify the part that will remain in service and the part that needs to be replaced. A photograph of the connection face or terminal markings helps resolve differences between versions that share the same housing.
For a new installation, describe what the instrument must measure or control and where the reading will be used. Give the normal operating conditions separately from short peaks. Include units with every numerical value. A drawing with the available space, orientation and access for removal gives the supplier a practical starting point.
Add the quantity for each position and identify any accessories that belong with it. Where a particular drawing, material certificate or calibration record is required, mention it with the line item. Submit the list with the destination country so that the response can cover the requested configuration and delivery.