Sierra Instruments SmartTrak 100
Sierra Instruments

SmartTrak 100 thermal mass flow controller and meter

Thermal mass meters & flow controllers

Manufacturer
Sierra Instruments
Model
SmartTrak 100
Instrument type
Thermal mass meters & flow controllers
Flow range
0–10 sccm up to 0–1000 slpm (0.1 Nml/min to 1300 Nl/min)
Accuracy
±0.5 % or ±1.0 % of full scale
Repeatability
±0.2 % of full scale
Gas selection
up to 10 gases in the instrument
Max. pressure
34.5 barg (500 psig)
Operating temperature
0 to 50 °C
Response time
time constant 300 ms; ≈2 s to within ±2 %
Outputs
4–20 mA, 0–20 mA, 0–5 / 1–5 / 0–10 V DC
Communication
RS-232, RS-485 Modbus RTU, Foundation Fieldbus, Profibus DP
Wetted materials
316 stainless steel; Viton, Neoprene, Kalrez seals
Connections
compression, VCO, VCR, female NPT
Power supply
24 V DC
Request a quote
Review enquiry

SmartTrak 100 is a series of digital thermal mass flow meters and mass flow controllers for gases. A meter measures and displays the mass flow; a controller additionally contains a proportional valve and keeps the flow at a setpoint given locally, by an analogue signal or over a digital interface. Because the measurement is mass-based, readings do not need correction for gas pressure and temperature.

The gas stream divides into a laminar flow element and a capillary sensor tube. Two platinum resistance windings on the capillary heat the gas; the temperature difference between them is proportional to the mass flow and is evaluated by the microprocessor. The laminar flow element keeps the split ratio constant, giving a linear signal over the whole range. The controller valve is a frictionless hovering design without sliding parts.

Full-scale ranges start at 0–10 sccm and reach 0–1000 slpm; metric versions cover 0.1 Nml/min to 1300 Nl/min. Accuracy is ±1 % of full scale as standard and ±0.5 % as an option, with repeatability of ±0.2 %. Each instrument is calibrated at ten points against primary standards. Up to ten gases can be stored and selected in the field, so one instrument can be switched between gases without recalibration. The time constant is 300 ms, and the controller reaches ±2 % of a new setpoint in about two seconds.

Bodies are made of 316 stainless steel with Viton, Neoprene or Kalrez seals for corrosive and reactive gases. Maximum pressure is 34.5 barg. Inlet and outlet fittings can be compression, VCO, VCR or female NPT. A pilot display module with keypad shows flow and totals and can also be mounted remotely. Outputs include 4–20 mA, 0–20 mA and 0–5, 1–5 or 0–10 V DC; digital communication is available over RS-232, RS-485 Modbus RTU, Foundation Fieldbus and Profibus DP. The instruments run on 24 V DC; power supplies, cables, a USB converter and remote display modules are offered as accessories. A 10 µm inline filter is recommended upstream.

Typical applications are semiconductor and vacuum processes, gas mixing and blending, engine test and emissions measurement, analytical instruments, life science and laboratory research. Please state gas, flow range, pressure and required interface with your enquiry.

Part numbers · 41

M100L, M100M, M100H, M100H1, M100H2, C100L, C100M, C100H, C100H1, C100H2, 100-T8D, 100-T8F, 100-T10D, 100-T10F, 100-Analog Cable, 100-CRN, 100-SerialUSB, 100-C0, 100-RDO, 100-CMNR, 100-CMDD, 100-CRC, 100-DCC, C25, C100, M100, C100 HP, M100 HP, C101, M101, C140, SmartTrak 100 Series, MassTrack 810C, TopTrack 822, TopTrack 826, FastFlo 620S, Steel-Mass 640S, FlatTrak 780S, Ultra High Purity 780S, QuadraTherm 780i, QuadraTherm 640i

Industry applications

Selection notes

State the minimum, normal and peak flow with units. Pipe bore, fluid properties and the available straight run affect the choice of measuring principle.

For an installed instrument, copy the complete ordering code and its suffixes from the nameplate. Identify the part that will remain in service and the part that needs to be replaced. A photograph of the connection face or terminal markings helps resolve differences between versions that share the same housing.

For a new installation, describe what the instrument must measure or control and where the reading will be used. Give the normal operating conditions separately from short peaks. Include units with every numerical value. A drawing with the available space, orientation and access for removal gives the supplier a practical starting point.

Add the quantity for each position and identify any accessories that belong with it. Where a particular drawing, material certificate or calibration record is required, mention it with the line item. Submit the list with the destination country so that the response can cover the requested configuration and delivery.