Gems Sensors & Controls UCL-510
Gems Sensors & Controls

UCL-510 ultrasonic continuous level transmitter for small tanks

Ultrasonic level transmitters

Manufacturer
Gems Sensors & Controls
Model
UCL-510
Instrument type
Ultrasonic level transmitters
Technology
ultrasonic, non-contact
Max. range
49.2 in (1250 mm)
Accuracy
0.125 in (approx. 3 mm)
Resolution
0.019 in (approx. 0.5 mm)
Output
4–20 mA
Loop fail-safe
4, 20, 21 or 22 mA, or hold last
Supply voltage
24 V DC
Max. loop resistance
400 Ω
Functions
switch, controller, transmitter; selectable hysteresis
Temperature compensation
automatic
Ambient temperature
−31 to 140 °F (approx. −35 to 60 °C)
Pressure
30 psi
Enclosure
Type 6P encapsulated, corrosion-resistant, submersible
Materials
PX/ABS FR, PVDF stem, Santoprene, Viton
Connections
1 in NPT; 48 in polyethylene cable
Request a quote
Review enquiry

UCL-510 is a non-contact ultrasonic continuous level transmitter for general purpose level measurement in small tanks. It has a maximum range of 49.2 inches (1250 mm) and is suited to corrosive and dirty liquids, since the sensor does not touch the medium and has no moving parts.

The sensor emits ultrasonic pulses towards the liquid surface and measures the time until the echo returns, from which the distance and thus the level are calculated. Automatic temperature compensation over the range corrects for the influence of air temperature on the speed of sound. Accuracy is 0.125 in (about 3 mm) and resolution 0.019 in (about 0.5 mm).

The UCL-510 is a multi-function unit that combines switch, controller and transmitter capabilities. Its 4–20 mA output transmits the level continuously, while switching functions with selectable hysteresis can be used for control or alarm. The loop fail-safe state can be set to 4 mA, 20 mA, 21 mA or 22 mA, or to hold the last value. The transmitter is supplied with 24 V DC, and the maximum loop resistance is 400 Ω.

The enclosure is a Type 6P encapsulated design that is corrosion-resistant and submersible. Materials of construction include flame-retardant PX/ABS, a PVDF stem and mounting, Santoprene and Viton. The process connection is a 1 in NPT thread, and the electrical connection a 48-inch polyethylene cable. The transmitter is rated for 30 psi and an ambient temperature range of −31 to 140 °F (about −35 to 60 °C).

Where surface turbulence or foam would disturb the measurement, the sensor can be mounted in a standpipe of 1 to 6 inches in diameter, which filters these effects. Part numbers 225100 and 227100 are listed for the series.

Stated applications are level monitoring of small tanks, corrosive and dirty media and standpipe installations. Please send the tank height, the liquid, the mounting arrangement and the required fail-safe setting with your enquiry.

Part numbers · 2

225100, 227100

Industry applications

Selection notes

Identify the vessel height, nozzle position and surface conditions. A point alarm and a continuous level reading require different outputs.

For an installed instrument, copy the complete ordering code and its suffixes from the nameplate. Identify the part that will remain in service and the part that needs to be replaced. A photograph of the connection face or terminal markings helps resolve differences between versions that share the same housing.

For a new installation, describe what the instrument must measure or control and where the reading will be used. Give the normal operating conditions separately from short peaks. Include units with every numerical value. A drawing with the available space, orientation and access for removal gives the supplier a practical starting point.

Add the quantity for each position and identify any accessories that belong with it. Where a particular drawing, material certificate or calibration record is required, mention it with the line item. Submit the list with the destination country so that the response can cover the requested configuration and delivery.

Related instruments

Find equipment

DEVICEEU