LUMEL VA701
LUMEL

VA701 portable voltage and current calibrator, 0–24 mA and 0–20 V

Multifunction process calibrators

Manufacturer
LUMEL
Model
VA701
Instrument type
Multifunction process calibrators
Current
0 to 24 mA DC, measure or source
Voltage
0 to 20 V DC, measure or source
Mode
measure or source, not simultaneously
Max voltage
30 V between jacks and earth
Temperature coefficient
±0.01 % of range per °C
Operating temperature
−10 to 50 °C
Storage temperature
−40 to 60 °C
Relative humidity
95 % to 30 °C; 75 % to 40 °C; 45 % to 50 °C; 35 % to 55 °C
Operating altitude
up to 2000 m
Power
6 × AAA batteries
Dimensions
204 × 99 × 46 mm
Weight
460 g
Approval
CE
Request a quote
Review enquiry

The VA701 is a portable process calibrator for electrical signals. It works both as a source and as a measuring instrument for DC current and DC voltage, the two signal types most widely used between field instruments, controllers and control systems, and is intended for calibration of industrial instruments and field maintenance.

In current mode the VA701 measures or outputs a 0 to 24 mA DC loop signal, which covers the standard 4–20 mA range with margin above and below. In voltage mode it measures or outputs 0 to 20 V DC. The instrument can either measure or source at a given time; measuring and sourcing simultaneously is not possible.

When sourcing, the calibrator can simulate a transmitter signal to check indicators, recorders, controller inputs and control system channels. When measuring, it reads the output of a transmitter or converter, so the instrument can be compared with its specified characteristic.

The temperature coefficient is ±0.01 % of range per °C within the specified temperature bands. The maximum voltage between the jacks and earth is 30 V. Permitted operating temperature is −10 to 50 °C and storage temperature −40 to 60 °C. Relative humidity limits are 95 % up to 30 °C, 75 % up to 40 °C, 45 % up to 50 °C and 35 % up to 55 °C, and the instrument may be used at altitudes up to 2000 m.

The calibrator is powered by six AAA batteries, which makes it independent of mains power in the field. It measures 204 × 99 × 46 mm and weighs 460 g. The device carries the CE mark.

Applications named are the calibration of industrial instruments and field maintenance. Please state the required quantity in your enquiry and whether batteries and test leads should be included, so that the quotation covers everything needed for immediate use.

Industry applications

Selection notes

Identify the instrument under test and the points to reproduce. State the required uncertainty, reference conditions and certificate format.

For an installed instrument, copy the complete ordering code and its suffixes from the nameplate. Identify the part that will remain in service and the part that needs to be replaced. A photograph of the connection face or terminal markings helps resolve differences between versions that share the same housing.

For a new installation, describe what the instrument must measure or control and where the reading will be used. Give the normal operating conditions separately from short peaks. Include units with every numerical value. A drawing with the available space, orientation and access for removal gives the supplier a practical starting point.

Add the quantity for each position and identify any accessories that belong with it. Where a particular drawing, material certificate or calibration record is required, mention it with the line item. Submit the list with the destination country so that the response can cover the requested configuration and delivery.

Related instruments