Surent G480
Flow

Variable area flowmeters
The VF series are direct reading variable area flowmeters with bodies precision machined from solid, clear acrylic blocks. They are suitable for both gas and liquid service and are offered with or without an integral precision metering valve. Because the body is completely transparent, the float and the flow can be inspected visually from several angles.
The flowmeter works on the variable area principle. The gas or liquid flows upwards through a precisely machined tapered bore and lifts a float; the higher the flow, the higher the float rises, until the annular gap around it is large enough for the flow to pass. The position of the float is read on the scale. The meters are mounted vertically with the inlet at the bottom and the outlet at the top. The precision machined tapered bore gives high repeatability.
Two scale lengths are available: VFA with a 2-inch (50.8 mm) scale and an accuracy of 5 % of full scale, and VFB with a 4-inch (101.6 mm) scale and an accuracy of 3 % of full scale. Direct reading scales are hot-stamped into the plastic, which removes the need for conversions; a white background improves the visibility of the float. Arbitrary scales are available as an option.
Without valve, the meters are rated for 100 psig (6.9 bar) at 65 °C or 150 psig (10 bar) at 38 °C; with valve, for 100 psig at 48 °C. Connections are 1/8-inch female NPT, with back or end connection options, and metal mounting inserts allow the meter to be supported directly by the system piping. Inlet piping should be at least as large as the connection. Wetted fittings are brass as standard and stainless steel as an option; floats are available in stainless steel, black glass, aluminium or K-Monel, and O-rings in Buna-N or fluoroelastomer. VFA meters weigh 0.11 to 0.14 kg and VFB meters 0.20 to 0.25 kg. The meters are RoHS compliant.
Options include end connections, an aluminium tag, an adjustable pointer flag, a PTFE valve seat, special cleaning, stainless steel wetted parts, fluoroelastomer O-rings, a version without logo and a NIST-traceable calibration certificate.
Listed applications are medical and laboratory equipment, air samplers, gas analysers, pollution monitors, chemical injectors and cabinet purging. Please send the medium, flow range, pressure and required options with your enquiry.
VFA-1, VFA-2, VFA-4, VFA-9, VFA-25, VFA-27, VFA-34, VFA-43, VFB-1, VFB-4, VFB-6, VFB-8, VFB-34, VFB-41, VFB-66, VFB-80, VFA-X, VFA-X-SS, VFA-X-BV***, VFA-X-SSV***, VFA-X-EC, VFA-X-EC-SS, VFB-X, VFB-X-SS, VFB-X-BV***, VFB-X-SSV***, VFB-X-EC, VFB-X-EC-SS
State the minimum, normal and peak flow with units. Pipe bore, fluid properties and the available straight run affect the choice of measuring principle.
For an installed instrument, copy the complete ordering code and its suffixes from the nameplate. Identify the part that will remain in service and the part that needs to be replaced. A photograph of the connection face or terminal markings helps resolve differences between versions that share the same housing.
For a new installation, describe what the instrument must measure or control and where the reading will be used. Give the normal operating conditions separately from short peaks. Include units with every numerical value. A drawing with the available space, orientation and access for removal gives the supplier a practical starting point.
Add the quantity for each position and identify any accessories that belong with it. Where a particular drawing, material certificate or calibration record is required, mention it with the line item. Submit the list with the destination country so that the response can cover the requested configuration and delivery.