EDT 101
Temperature

Temperature transmitters
The VP5 series are measuring transducers that convert the signal of a Pt100 temperature sensor into a standardised voltage or current output. They are compact DIN-rail devices for harsh environments, machine controllers and marine applications and carry DNV, LR and BV approvals as well as the CE mark.
The transducer evaluates the resistance of a two-wire Pt100 sensor according to EN 60751 class B and converts it into a linear output signal. The sensor input is galvanically isolated from the operating voltage. The measuring range can be fine-adjusted with a sealed trimming potentiometer. The transducer monitors the sensor circuit for short circuit and broken wire, and an integrated LED shows the operating status.
Four measuring ranges are available: 0–120 °C (VP51), 0–150 °C (VP52), 0–200 °C (VP53) and −30–120 °C (VP511); the overall measuring range is −30 to +200 °C. Each range is offered with 0–10 V (-G1), 2–10 V (-G2), 0–20 mA (-I1) or 4–20 mA (-I2) output, for example VP51-I2 for 0–120 °C and 4–20 mA. Voltage outputs are short-circuit proof with a maximum load current of 20 mA; the maximum load resistance for current outputs is 500 Ω. The linearity error is less than ±0.3 %, and the error class is 1.5 % to IEC 51-1. Pt1000 sensors are mentioned as an option.
The transducer is supplied with 9 to 32 V DC and draws about 50 mA at 24 V DC. It works at ambient temperatures from −20 °C to +70 °C and can be stored at −45 °C to +85 °C, with humidity up to 96 % RH. Vibration resistance is 15 g at 10–100 Hz to IEC 60068-2-6 and shock resistance 300 m/s² with 18 ms dwell time to IEC 60068-2-27. The body is made of green thermoplastic polyester, flame-inhibiting and self-extinguishing (V0). Connections are gold-plated A6.3 × 0.8 tab connectors to DIN EN 46244, and the device snaps onto TS32 G-rails or TS35 top-hat rails in any position, side by side. It weighs 55 g. EMC tests cover IEC 61000-4-2 to -4-6 and DIN EN 61000-6-2/-6-4; DIN EN 50155 and IEC 60553 are also listed.
Listed fields of use are marine applications, harsh environments and machine controllers.
Please state the measuring range, output signal and quantity with your enquiry.
VP51-I2, VP51-G1, VP51-G2, VP52-G1, VP52-G2, VP53-G1, VP53-G2, VP511-G1, VP511-G2, VP511-I1, VP511-I2, VP51-I1, VP52-I1, VP52-I2, VP53-I1, VP53-I2, VPT511-G1, VPT511-G2, VPT511-I1, VPT511-I2, VPT51-I1, VPT51-I2, VPT52-I1, VPT52-I2, VPT53-I1, VPT53-I2
Match the sensing element, insertion length and wiring to the existing input. Separate the process temperature from the ambient temperature at the connection head.
For an installed instrument, copy the complete ordering code and its suffixes from the nameplate. Identify the part that will remain in service and the part that needs to be replaced. A photograph of the connection face or terminal markings helps resolve differences between versions that share the same housing.
For a new installation, describe what the instrument must measure or control and where the reading will be used. Give the normal operating conditions separately from short peaks. Include units with every numerical value. A drawing with the available space, orientation and access for removal gives the supplier a practical starting point.
Add the quantity for each position and identify any accessories that belong with it. Where a particular drawing, material certificate or calibration record is required, mention it with the line item. Submit the list with the destination country so that the response can cover the requested configuration and delivery.