EDT 96
Pressure

Pressure gauges
The 477AV is a handheld digital manometer for heating, ventilation and air conditioning work. It measures positive, negative and differential pressure of air and compatible gases and, connected to a Pitot tube, calculates air velocity and volumetric air flow directly, so a technician can carry out duct traverses and air balancing with one instrument.
The heart of the instrument is a differential pressure sensor. The series includes models with different pressure ranges. Accuracy is ±0.5 % of full scale at normal room temperature (15.6 to 25.6 °C) and ±1.5 % of full scale from 0 to 15.6 °C and from 25.6 to 40 °C; pressure hysteresis is ±0.1 %. Nine English and metric engineering units can be selected.
Velocity and flow are calculated by the instrument from the measured velocity pressure. Up to 40 readings can be stored and recalled together with minimum, maximum and average values, which is convenient for traverse measurements across a duct cross-section. The 4-digit LCD has a backlight for dark technical rooms and ceiling voids.
The electronics are housed in a rugged aluminium case that protects the instrument during transport and site work; a magnetic rubber boot is available as an accessory. Pressure connections are two barbed fittings for 1/8 in or 3/16 in ID tubing, or compression fittings for 1/8 in ID × 1/4 in OD tube, depending on the model. Power comes from a single 9 V alkaline battery. The instrument is CE marked and RoHS compliant.
The -AV version is supplied as an all-in-one air velocity kit; a carrying case is available. Options include a certificate of calibration, factory calibration and a traceable calibration certificate. Typical tasks are airflow monitoring, duct static pressure checks, Pitot tube traverses, commercial air balancing, building-zone pressure checks and verification of pressure instruments by maintenance staff. Please state the required pressure range and whether you need the air velocity kit.
477AV-000, 477AV-00, 477AV-0, 477AV-1, 477AV-2, 477AV-3, 477AV-4, 477AV-5, 477AV-6, 477AV-7, 477AV-8, 477AV-1-NIST, 477AV-4-NIST, 477AV-000T, 477AV-00T, A-47X-BOOT, A-402A, A-201, A-303, A-397, A-463, A-375, 166T, A-472T
Specify whether the reference is gauge, absolute or differential pressure. Include the normal span and the largest pressure the instrument may see.
For an installed instrument, copy the complete ordering code and its suffixes from the nameplate. Identify the part that will remain in service and the part that needs to be replaced. A photograph of the connection face or terminal markings helps resolve differences between versions that share the same housing.
For a new installation, describe what the instrument must measure or control and where the reading will be used. Give the normal operating conditions separately from short peaks. Include units with every numerical value. A drawing with the available space, orientation and access for removal gives the supplier a practical starting point.
Add the quantity for each position and identify any accessories that belong with it. Where a particular drawing, material certificate or calibration record is required, mention it with the line item. Submit the list with the destination country so that the response can cover the requested configuration and delivery.