Type 8203
Liquid analysis

pH & ORP electrodes
The 6560S-10C is a combination ion selective electrode (ISE) for the measurement of chloride ions in liquid samples. Measuring and reference functions are combined in one body, so a single electrode connected to a suitable meter is enough to determine chloride concentration.
The electrode uses an Ag/AgCl internal electrode and a movable sleeve liquid junction. In the sample solution it generates a potential difference that depends on the chloride ion activity; the slope is −53 to −62 mV per decade of concentration. The internal filling solution is 1 mol/L KNO₃.
The measuring range is 0.35 to 35,000 mg/L Cl⁻, corresponding to 1 × 10⁻⁵ to 1 mol/L, with an accuracy of ±2 %. The electrode can be used at sample temperatures from 0 to 50 °C and within a pH range of 3 to 11 (at 350 mg/L Cl⁻).
Wetted parts are made of PVC, PSU and AgCl. The electrode is 150 mm long with a diameter of 16.5 mm and is fitted with a 1 m cable and a BNC connector. It complies with RoHS according to EN 50581 and with the REACH regulation; GB/T 26572 is also listed.
Matching consumables and spare parts are available: the replaceable electrode tip 7660S, the chloride electrode filling solution 500-CL-IFS (500 mL), the standard solutions 500-CL-SL (100 mg/L chloride, 500 mL) and 500-CL-SH (1000 mg/L chloride, 500 mL), and the chloride ionic strength adjustor 500-CL-ISA (500 mL), which is added to samples and standards.
The electrode is listed for a broad range of samples: agriculture, river and tap water, plant tissue, soils, boiler feed water, clinical analysis, sweat and urine, cement, plating baths, and dairy, food and beverage samples.
When enquiring, please state the number of electrodes and which consumables, such as replacement tips, filling solution, standards or ionic strength adjustor, should be included, together with the meter to which the electrode will be connected.
6560S-10C, 7660S, 500-CL-IFS, 500-CL-SL, 500-CL-SH, 500-CL-ISA
Record the sample temperature, expected concentration and how the probe is mounted. Identify the mating transmitter when replacing a sensor.
For an installed instrument, copy the complete ordering code and its suffixes from the nameplate. Identify the part that will remain in service and the part that needs to be replaced. A photograph of the connection face or terminal markings helps resolve differences between versions that share the same housing.
For a new installation, describe what the instrument must measure or control and where the reading will be used. Give the normal operating conditions separately from short peaks. Include units with every numerical value. A drawing with the available space, orientation and access for removal gives the supplier a practical starting point.
Add the quantity for each position and identify any accessories that belong with it. Where a particular drawing, material certificate or calibration record is required, mention it with the line item. Submit the list with the destination country so that the response can cover the requested configuration and delivery.