PACE6000-ADC
Calibration

Temperature calibrators
The 9132 and 9133 are portable infrared calibrators with a blackbody target. They provide a stable source of thermal radiation with known temperature and emissivity, against which infrared thermometers and thermal imagers are checked and calibrated, in the laboratory and in the field. The two models complement each other: the 9132 covers high temperatures from 50 °C to 500 °C, the 9133 covers low and ambient temperatures from −30 °C to 150 °C.
The target is a flat surface 57 mm in diameter with a high emissivity of 0.95 in the 8–14 µm band used by most industrial infrared instruments. It is heated or cooled to the set-point by a controller with a platinum control sensor. Accuracy of the 9132 is ±0.5 °C at 100 °C and ±0.8 °C at 500 °C, with stability of ±0.1 °C and ±0.3 °C respectively; the 9133 reaches ±0.4 °C accuracy and ±0.1 °C stability. The display resolution is 0.1°.
Both instruments are fast enough for practical use: the 9132 heats from 50 to 500 °C in about 30 minutes and cools from 500 to 100 °C in the same time; the 9133 heats from 25 to 150 °C in 15 minutes and cools from 25 to −20 °C in 15 minutes. Up to eight set-points can be stored, and the scan rate between them is adjustable from 0.1 to 99.9 °C per minute, so a sequence of calibration points can be run quickly. A high-limit parameter protects the instrument.
Operation is done on the front panel or remotely via the RS-232 interface with a DB-9 connector at baud rates up to 9600, with an IEEE-488 interface as an option, which allows integration into automated calibration systems. Calibration constants are stored in the instrument. The 9132 weighs 1.8 kg and measures 102 × 152 × 178 mm; the 9133 weighs 4.6 kg and measures 152 × 286 × 267 mm. Both run on 115 V or 230 V AC.
Typical uses are field and laboratory calibration of IR thermometers, verification of thermal imagers, quality control of non-contact temperature measurement and routine instrument checks. A shielded serial cable and a report of calibration are available. Please specify which model, or both, and the temperatures you need to cover.
9132, 9133
Identify the instrument under test and the points to reproduce. State the required uncertainty, reference conditions and certificate format.
For an installed instrument, copy the complete ordering code and its suffixes from the nameplate. Identify the part that will remain in service and the part that needs to be replaced. A photograph of the connection face or terminal markings helps resolve differences between versions that share the same housing.
For a new installation, describe what the instrument must measure or control and where the reading will be used. Give the normal operating conditions separately from short peaks. Include units with every numerical value. A drawing with the available space, orientation and access for removal gives the supplier a practical starting point.
Add the quantity for each position and identify any accessories that belong with it. Where a particular drawing, material certificate or calibration record is required, mention it with the line item. Submit the list with the destination country so that the response can cover the requested configuration and delivery.