HORIBA APOA-370
HORIBA

APOA-370 ambient ozone monitor with cross-flow modulated UV absorption

Process gas analysers & chromatographs

Manufacturer
HORIBA
Model
APOA-370
Instrument type
Process gas analysers & chromatographs
Measured component
ozone (O3) in ambient air
Measuring principle
cross-flow modulated non-dispersive UV absorption (NDUV)
Reference gas
heated de-ozoniser removes O3
Wetted materials
glass or PTFE only
Standard ranges
0–0.1 / 0.2 / 0.5 / 1.0 ppm; auto or manual, remote switching
Optional ranges
4 ranges up to 0–10 ppm, range ratio within 10
Repeatability / linearity
±1.0 % FS each
Sample flow
approx. 0.7 l/min
Analogue outputs
0–1 V, 0–10 V or 4–20 mA (instantaneous and integrated, or moving average)
Digital I/O
contact input/output, RS-232C
Power supply
100 to 240 V AC, 50/60 Hz (to be specified)
Dimensions, mass
430 × 550 × 221 mm (W × D × H), approx. 20 kg
Certification
U.S. EPA reference method EQOA-0506-160; RoHS, REACH, EMC
Request a quote
Review enquiry

APOA-370 is an ambient ozone monitor for continuous monitoring of atmospheric ozone concentration. It belongs to the AP-370 series and is also used for monitoring airborne molecular contamination in clean rooms of the electronics industry.

The analyser uses a cross-flow modulated ultraviolet absorption method (NDUV). Sample gas and reference gas are alternately passed through the detector; the reference gas is produced by a heated de-ozoniser that removes ozone, so the difference in UV absorption between the two gives the ozone concentration. An independent, internal dry-method sampling device analyses the gas in its unaltered state, and only inert materials, glass or PTFE, come into contact with the sample.

Standard ranges are 0–0.1, 0.2, 0.5 and 1.0 ppm. As an option, four ranges can be selected up to 0–10 ppm within a range ratio of 10. In both cases auto-range or manual range selection is possible, and ranges can be switched remotely. Repeatability and linearity are each ±1.0 % of full scale, and the sample flow is approximately 0.7 l/min. The analyser is designated as U.S. EPA reference method EQOA-0506-160.

Analogue outputs of 0–1 V, 0–10 V or 4–20 mA are provided, to be specified, in two systems: either the instantaneous value and the integrated value, or the moving average value. Contact inputs and outputs and an RS-232C interface are also available. The analyser is supplied for 100, 110, 115, 120, 220, 230 or 240 V AC, 50/60 Hz, to be specified with the order.

The unit measures 430 × 550 × 221 mm (W × D × H) and weighs approximately 20 kg. It complies with RoHS, REACH and EMC requirements. The OZGU-370SE ozone transfer standard is available as an accessory for checking the monitor.

Listed applications include environmental ambient air monitoring, monitoring of airborne molecular contamination in electronics clean rooms, semiconductor manufacturing processes, automotive manufacturing, drinking water utilities and research and testing laboratories.

Please state the required range, output signals and supply voltage, and whether the OZGU-370SE transfer standard is needed.

Part numbers · 1

OZGU-370SE

Industry applications

Selection notes

Identify the target gas, background gas and concentration units. State whether the sampling system is already installed or belongs in the enquiry.

For an installed instrument, copy the complete ordering code and its suffixes from the nameplate. Identify the part that will remain in service and the part that needs to be replaced. A photograph of the connection face or terminal markings helps resolve differences between versions that share the same housing.

For a new installation, describe what the instrument must measure or control and where the reading will be used. Give the normal operating conditions separately from short peaks. Include units with every numerical value. A drawing with the available space, orientation and access for removal gives the supplier a practical starting point.

Add the quantity for each position and identify any accessories that belong with it. Where a particular drawing, material certificate or calibration record is required, mention it with the line item. Submit the list with the destination country so that the response can cover the requested configuration and delivery.

Related instruments