HORIBA APOA-380WM
HORIBA

APOA-380WM wall-mount ozone monitor with NDUV

Process gas analysers & chromatographs

Manufacturer
HORIBA
Model
APOA-380WM
Instrument type
Process gas analysers & chromatographs
Measured component
ozone (O3)
Measuring principle
cross-modulation NDUV, UV LED source
Measuring ranges
0–0.1 / 0.2 / 0.5 / 1.0 / 2.0 / 5.0 / 10 ppm
Lower detection limit (2σ)
< 0.3 ppb
Repeatability / linearity
±1.0 % of full scale each
Zero / span drift
< 0.5 ppb / < 0.5 % FS per 24 h
Response time (T95)
< 60 s
Sample flow
approx. 0.6 l/min, internal pump
Communication
Ethernet (Modbus TCP) or RS-232C (Modbus RTU); USB flash drive
Analogue outputs (option)
up to 3 insulated: DC 4–20 mA, 0–0.1, 0–1, 0–5 or 0–10 V
Digital I/O (option)
relay contact outputs (3 range + 6 other); up to 4 non-isolated inputs
Display
7-inch colour LCD with touch panel
Power supply
AC 100–240 V ±10 %, 50/60 Hz, 60 W
Operating conditions
0–40 °C, ≤ 85 % RH
Dimensions, weight
424 × 251 × 484 mm (W × D × H), approx. 19 kg
Request a quote
Review enquiry

APOA-380WM is the wall-mount type of the APOA-380 ozone monitor. It measures ozone (O3) in ambient air continuously and online, and is designed for air quality monitoring, for detecting ozone leakage in drinking water and wastewater treatment facilities, for occupational health and safety and for industrial applications.

The analyser uses cross-modulation non-dispersive ultraviolet absorption (NDUV) with a UV LED as the light source. A heated ozone decomposer serves as a reference gas purifier and minimises interference from changes in moisture. An internal sampling pump draws the sample through the analyser at approximately 0.6 l/min.

The measuring range can be selected from up to seven ranges between 0–0.1 and 0–10 ppm (0.1, 0.2, 0.5, 1.0, 2.0, 5.0 and 10 ppm). Results can be displayed in ppb, µg/m³, ppm or mg/m³. The lower detection limit (2σ) is below 0.3 ppb, repeatability and linearity are ±1.0 % of full scale, zero drift is below 0.5 ppb and span drift below 0.5 % of full scale per 24 hours, and the response time T95 is under 60 seconds.

The monitor is operated via a 7-inch colour LCD touch screen and can also be operated remotely from a PC, tablet or smartphone. Communication is via one Ethernet port with Modbus TCP or one RS-232C port with Modbus RTU, plus a USB flash drive port. As options, up to three insulated analogue outputs (DC 4–20 mA, 0–0.1 V, 0–1 V, 0–5 V or 0–10 V), relay contact outputs for range indication (three channels) and other signals (six channels), and up to four non-isolated digital inputs can be added; a wireless connection requires a separate device. Power supply is AC 100–240 V ±10 %, 50/60 Hz, 60 W.

The wall-mount housing measures 424 × 251 × 484 mm (W × D × H) and the unit weighs approximately 19 kg. It is designed for operation at 0–40 °C and up to 85 % relative humidity. The modular design allows fast replacement in the field, and remote diagnostics support maintenance. The product complies with RoHS, REACH and EMC requirements.

Please include the required measuring range, the communication interface and any analogue or relay outputs needed when you send your enquiry.

Part numbers · 1

APOA-380WM

Industry applications

Selection notes

Identify the target gas, background gas and concentration units. State whether the sampling system is already installed or belongs in the enquiry.

For an installed instrument, copy the complete ordering code and its suffixes from the nameplate. Identify the part that will remain in service and the part that needs to be replaced. A photograph of the connection face or terminal markings helps resolve differences between versions that share the same housing.

For a new installation, describe what the instrument must measure or control and where the reading will be used. Give the normal operating conditions separately from short peaks. Include units with every numerical value. A drawing with the available space, orientation and access for removal gives the supplier a practical starting point.

Add the quantity for each position and identify any accessories that belong with it. Where a particular drawing, material certificate or calibration record is required, mention it with the line item. Submit the list with the destination country so that the response can cover the requested configuration and delivery.

Related instruments

Find equipment

DEVICEEU