Sierra Instruments BoilerTrak 620S-BT
Sierra Instruments

BoilerTrak 620S-BT insertion thermal mass flow meter for natural gas and propane

Thermal mass meters & flow controllers

Manufacturer
Sierra Instruments
Model
BoilerTrak 620S-BT
Instrument type
Thermal mass meters & flow controllers
Measuring principle
thermal dispersion, direct mass flow
Accuracy
±1 % of full scale
Repeatability
±0.2 % of full scale
Turndown
100:1
Maximum velocity
17,000 sfpm (18–24 V supply); 12,000 sfpm (15–18 V supply)
Response time
200 ms
Gas temperature
−40 to 80 °C (−40 to 176 °F)
Ambient temperature
−40 to 50 °C (−40 to 120 °F)
Pipe sizes
optimised for 1, 2, 3 and 4 inch Sch. 40
Probe
3/8 in (9.5 mm) diameter, lengths 3.7 in (9.3 cm) to 24 in (61 cm)
Outputs
linear 0–5 VDC and 4–20 mA, programmable pulse
Interfaces
RS-232, Modbus RTU
Power supply
15–18 VDC, max. 625 mA, or 18–24 VDC, max. 850 mA (regulated)
Enclosure rating
NEMA 4X, IP65
Approvals
CE, CRN, RoHS 3, REACH, WEEE
Request a quote
Review enquiry

The BoilerTrak 620S-BT is an insertion thermal mass flow meter intended for natural gas and propane measurement on boilers and heaters. It is offered as a metering solution for boiler and heater efficiency work, stationary combustion fuel metering and greenhouse gas reporting, and it is certified for GHG measurement to EPA 40 CFR Part 98 and for Boiler MACT compliance of area source and major source boilers (40 CFR 63, subparts JJJJJJ and DDDDD).

The meter works on the thermal dispersion principle and measures mass flow directly, so no separate pressure or temperature compensation is needed. The sensor has no moving parts, creates a low pressure drop and offers a high turndown of 100:1. It reacts to changes in flow rate within 200 milliseconds. Besides natural gas and propane, the media listed for the technology include methane, hydrogen, digester gas, inert gases and other clean, non-condensing gases.

Accuracy is ±1 % of full scale with a repeatability of ±0.2 % of full scale. The maximum gas velocity is 17,000 sfpm on 24 V models and 12,000 sfpm on 15 V models. The meter handles gas temperatures from −40 to 80 °C and ambient temperatures from −40 to 50 °C. It is optimised for 1, 2, 3 and 4 inch Schedule 40 pipe, with a small 3/8 inch (9.5 mm) sensor probe in lengths from 3.7 in (9.3 cm) to 24 in (61 cm).

Linear 0–5 VDC and 4–20 mA output signals are provided, together with a programmable pulse output for remote totalisation. RS-232 and Modbus RTU are available for digital communication. Two power variants exist: 15 V models operate on a regulated 15–18 VDC supply and draw up to 625 mA, while 24 V models operate on 18–24 VDC and draw up to 850 mA.

The enclosure is rated NEMA 4X and IP65, and remote enclosure mounting is available. The meter carries CE marking, CRN approvals for several Canadian provinces and conforms to RoHS 3, REACH and WEEE; ARRA compliance is also listed. Production is under ISO 9001:2015.

The insertion design allows easy installation in the field or retrofit into existing lines. Field flow-range adjustment, meter validation and diagnostics are supported through the product software. When enquiring, please state the gas, pipe size, flow range, required supply voltage and output signals.

Part numbers · 13

BoilerTrak 620S-BT, S-Series Pulse Output Information Sheet, 600/700 Series Product Software SIP134, 47-0493-D, CRN Approval, CRN Approval Ontario, CRN Approval Manitoba, QA-105-F, CRN Approval Quebec, CRN Approval British Columbia, CRN Approval Saskatchewan, Warranty Statement, Terms and Conditions

Industry applications

Selection notes

State the minimum, normal and peak flow with units. Pipe bore, fluid properties and the available straight run affect the choice of measuring principle.

For an installed instrument, copy the complete ordering code and its suffixes from the nameplate. Identify the part that will remain in service and the part that needs to be replaced. A photograph of the connection face or terminal markings helps resolve differences between versions that share the same housing.

For a new installation, describe what the instrument must measure or control and where the reading will be used. Give the normal operating conditions separately from short peaks. Include units with every numerical value. A drawing with the available space, orientation and access for removal gives the supplier a practical starting point.

Add the quantity for each position and identify any accessories that belong with it. Where a particular drawing, material certificate or calibration record is required, mention it with the line item. Submit the list with the destination country so that the response can cover the requested configuration and delivery.

Related instruments

Find equipment

DEVICEEU