Type 8203
Liquid analysis

Dissolved oxygen, chlorine & ozone sensors
DOTX45 is a two-wire, isolated dissolved oxygen transmitter system for continuous online monitoring of dissolved oxygen in solution. It works together with the DOE-45PA sensor, which has a polymer membrane-covered galvanic sensing element, and provides a 4–20 mA signal for a control system or recorder.
The system measures dissolved oxygen concentration and percent saturation using the membrane-covered galvanic principle with three electrodes (amperometric measurement). Oxygen diffuses through the membrane and produces a current proportional to its partial pressure. Because the permeability of the membrane depends on temperature, the transmitter compensates automatically with a Pt100 or Pt1000 element. The microprocessor-based electronics include a diagnostic that detects a punctured or torn membrane, and the sensor cartridge is replaceable.
The measuring range is 0 to 40 ppm or 0 to 200 % saturation. Repeatability is 0.1 % of span or better, sensitivity 0.05 % of span and non-linearity ±0.1 %. Stability is 0.1 % of span per 24 hours, non-cumulative, and the temperature drift of span or zero is 0.02 % of span per °C. The sensor reaches 90 % of a step change in less than three minutes; the transmitter itself responds within four seconds and warms up in four seconds.
The standard DOTX-45 is loop-powered from 16 to 35 V DC and fully isolated; the signal can be transmitted over 305 m. The DOTX-45-115VAC-R2 version is powered from 115 V AC and adds relays and dual 4–20 mA outputs. A large dual display shows the measured values. The transmitter operates from −20 to 60 °C; the process temperature range is −5 to 55 °C, the pressure rating 0 to 150 psig at 25 °C and the minimum sample flow 6 mm per second.
The enclosure is rated NEMA 4X and IP66 and is made of polycarbonate with stainless steel hardware; sensor materials include 316 SS, FKM and Noryl. Two PG9 openings and one 1-inch NPT centre opening are provided. The transmitter can be mounted on a wall, panel, pipe or DIN rail, or integrally with the sensor, and weighs 0.45 kg. EMC conformity follows EN 61326-1; the device is RoHS compliant.
Options include submersible sensing modules, replacement lead electrodes and unregulated power supplies. Please state the required power version, the mounting method and whether a submersible sensor is needed with your enquiry.
DOTX-45, DOTX-45-115VAC-R2
Record the sample temperature, expected concentration and how the probe is mounted. Identify the mating transmitter when replacing a sensor.
For an installed instrument, copy the complete ordering code and its suffixes from the nameplate. Identify the part that will remain in service and the part that needs to be replaced. A photograph of the connection face or terminal markings helps resolve differences between versions that share the same housing.
For a new installation, describe what the instrument must measure or control and where the reading will be used. Give the normal operating conditions separately from short peaks. Include units with every numerical value. A drawing with the available space, orientation and access for removal gives the supplier a practical starting point.
Add the quantity for each position and identify any accessories that belong with it. Where a particular drawing, material certificate or calibration record is required, mention it with the line item. Submit the list with the destination country so that the response can cover the requested configuration and delivery.