EDT 96
Pressure

Pressure gauges
The DPG210 is a high-accuracy digital pressure gauge with a 5-digit LCD and an additional measurement of ambient temperature. It is intended for precise pressure monitoring and for field calibration tasks, and is used for field calibration, leak testing, medical device testing and the monitoring of pressure systems.
The gauge measures pressure with a digital sensor and an accuracy of 0.05 % of full scale; long-term stability is ±0.2 % FS per year. The reading can be shown in ten selectable units, including MPa, psi and bar. Ambient temperature is displayed in °C or °F. The sampling rate is adjustable from 1 to 10 readings per second.
The series covers a wide choice of ranges; catalogue models from DPG210-005 to DPG210-500 are listed, and ranges up to 10,000 psi as well as metric ranges are available. Overload capability depends on the range: proof pressure is three times the rated pressure from 0 to 100 psi and twice the rated pressure from 200 to 5000 psi, while burst pressure is three times full scale. Process temperature is 0 to 40 °C, operating temperature −10 to 50 °C and storage up to 60 °C.
The housing is 304 stainless steel with IP65 protection, and the process connection is 1/4″ NPT male. The gauge runs on three AA alkaline batteries with a battery life of up to 3600 hours, or from a USB 2.0 port. Communication settings with address bit and baud rate are available as options. The gauge is CE, UKCA and RoHS compliant, and NIST-traceable calibration certificates can be supplied.
Calibration is adjusted at the factory; no field adjustment is provided. After installation, the zero should be cleared if mounting stress affects the reading.
For field calibration, the gauge is also offered as a complete kit with a pneumatic or hydraulic hand pump. Please state the pressure range, units and whether a calibration certificate is required in your enquiry.
DPG210-005, DPG210-015, DPG210-030, DPG210-060, DPG210-100, DPG210-200, DPG210-300, DPG210-500, DPG210-1.0KG, DPG210-10KG
Specify whether the reference is gauge, absolute or differential pressure. Include the normal span and the largest pressure the instrument may see.
For an installed instrument, copy the complete ordering code and its suffixes from the nameplate. Identify the part that will remain in service and the part that needs to be replaced. A photograph of the connection face or terminal markings helps resolve differences between versions that share the same housing.
For a new installation, describe what the instrument must measure or control and where the reading will be used. Give the normal operating conditions separately from short peaks. Include units with every numerical value. A drawing with the available space, orientation and access for removal gives the supplier a practical starting point.
Add the quantity for each position and identify any accessories that belong with it. Where a particular drawing, material certificate or calibration record is required, mention it with the line item. Submit the list with the destination country so that the response can cover the requested configuration and delivery.