BAUR DTL C
BAUR

DTL C automatic dissipation factor tester for insulating oils

Power equipment testing

Manufacturer
BAUR
Model
DTL C
Instrument type
Power equipment testing
Measured quantities
dissipation factor (tan δ), specific resistance, relative permittivity
Dissipation factor accuracy
up to 1 × 10⁻⁶
Specific resistance
up to 100 TΩm
Measuring voltage
positive or negative polarity
Test sequences
twelve, according to international standards
User procedures
up to ten programmable
Cell heating
rapid induction heating, precise temperature control
Cell draining
automatic, via magnetic discharge valve
Sample exchange
contactless, at operating temperature
Calibration
automatic calibration of the empty cell
Operation
fully automated
Request a quote
Review enquiry

DTL C is an automated diagnostic device for the condition evaluation of insulating oils used in electrical equipment. It determines the dielectric properties of an oil sample, which indicate how far the liquid has been affected by contamination and ageing, and therefore supports decisions on the further operation of the equipment filled with it.

The instrument measures three quantities: the dissipation factor (tan δ), the specific resistance and the relative permittivity. The dissipation factor describes the dielectric losses of the oil in an alternating field, while the specific resistance shows its insulating capability; it is measured with a positive or negative measuring voltage. Changes in these values reveal contamination, oxidation products and the effects of internal partial discharges.

The dissipation factor is measured with an accuracy of up to 1 × 10⁻⁶, and the specific resistance can be determined up to 100 TΩm. To bring the sample to the test temperature quickly, the measuring cell is heated by induction, and the temperature is regulated precisely during the measurement.

Operation is fully automated. The device contains twelve different test sequences according to international standards, and up to ten individual measuring procedures can be programmed by the user for in-house methods. After the measurement the cell is drained automatically through a magnetic discharge valve. The sample can be exchanged without contact while the cell remains at operating temperature, so that consecutive samples can be tested without waiting for the cell to cool down. The empty cell is calibrated automatically before the sample is measured.

Because heating, measurement, draining and empty-cell calibration all run automatically, a complete test sequence is carried out by the device itself once the sample has been filled in and the sequence has been selected.

Listed fields of use are monitoring the safe operation of cable networks, research and development, and the detection of contamination and oxidation in insulating liquids.

Please describe the type of insulating liquid, the test standards you need to follow and the expected number of samples in your enquiry, so that the configuration of the device can be clarified.

Part numbers · 1

DTL C

Industry applications

Selection notes

Match the instrument to the circuit and measurement category. Include the wiring arrangement and ratios of any external current or voltage transformers.

For an installed instrument, copy the complete ordering code and its suffixes from the nameplate. Identify the part that will remain in service and the part that needs to be replaced. A photograph of the connection face or terminal markings helps resolve differences between versions that share the same housing.

For a new installation, describe what the instrument must measure or control and where the reading will be used. Give the normal operating conditions separately from short peaks. Include units with every numerical value. A drawing with the available space, orientation and access for removal gives the supplier a practical starting point.

Add the quantity for each position and identify any accessories that belong with it. Where a particular drawing, material certificate or calibration record is required, mention it with the line item. Submit the list with the destination country so that the response can cover the requested configuration and delivery.

Related instruments