Magnetrol Eclipse 706
Magnetrol

Eclipse 706 loop-powered guided wave radar level transmitter

Radar level transmitters

Manufacturer
Magnetrol
Model
Eclipse 706
Instrument type
Radar level transmitters
Principle
guided wave radar, TDR
Supply
24 V DC, loop powered
Min dielectric constant
εr 1.4
Measuring range
12 in (30 cm) to 224 in
Outputs
4–20 mA; level, volumetric and flow output
Communication
HART, Modbus, PROFIBUS PA, Foundation fieldbus; FDT/DTM, DD
Display
graphic LCD
Enclosure
angled, dual compartment, NEMA 4X
Probes
coaxial; twin rod or cable; single rod or cable
Remote mounting
up to 3.6 m (12 ft)
Steam compensation
mechanical target on probe
Safety
SIL 2 / SIL 3
Approvals
ATEX, CSA, FM, TÜV, EAC, INMETRO, LR
Standards
EN 12952-11, EN 12953-9, ASME B31.1, B31.3, WHG § 63
Request a quote
Review enquiry

The Eclipse 706 is a high-performance, loop-powered 24 V DC guided wave radar (GWR) level transmitter for accurate level measurement in the process industries. It is used with various probe configurations to measure liquid levels in diverse industrial applications, including saturated steam environments.

The transmitter works on the time domain reflectometry (TDR) principle. Electromagnetic pulses are sent down a waveguide, the probe, and are reflected at the liquid surface; the transit time of the reflected pulse is measured to determine the level. Media with a dielectric constant as low as 1.4 can be measured. In steam applications, changes in the dielectric constant of the steam phase are compensated using a mechanical target on the probe.

Measuring range extends from a minimum of 12 in (30 cm) to a maximum of 224 in, depending on the probe. Besides level, the transmitter can provide volumetric and flow outputs. The 4–20 mA loop signal is complemented by HART, Modbus, PROFIBUS PA or Foundation fieldbus communication, and device integration is supported by FDT/DTM and DD. A graphic LCD shows values and menus.

The enclosure is angled with two compartments and rated NEMA 4X. Probe types are coaxial, twin element (rod or cable) and single element (rod or cable). The electronics can be mounted remotely up to 12 ft (3.6 m) from the probe. Process connections include flanged and threaded versions, with proprietary and standard ASME flanges; an adapter adds 3.5 inches to the installed height.

The transmitter is certified for SIL 2 and SIL 3 safety loops. Approvals include ATEX, CSA, FM, TÜV, EAC, INMETRO and LR, and boiler and piping references include EN 12952-11, EN 12953-9, ASME B31.1 and B31.3, as well as overfill protection to WHG § 63.

Applications named are power generation (steam drums and boilers), chemical and petrochemical industries, petroleum refining and pulp and paper. Local upgrades and retrofitting of legacy Model 705 probes are possible with the Eclipse PLUS+ adapter. Please send the medium, pressure, temperature, measuring range and required protocol with your enquiry.

Part numbers · 3

706-511A-310, 7AG-4300-A10-00-021, F21-4A2D-014

Industry applications

Selection notes

Identify the vessel height, nozzle position and surface conditions. A point alarm and a continuous level reading require different outputs.

For an installed instrument, copy the complete ordering code and its suffixes from the nameplate. Identify the part that will remain in service and the part that needs to be replaced. A photograph of the connection face or terminal markings helps resolve differences between versions that share the same housing.

For a new installation, describe what the instrument must measure or control and where the reading will be used. Give the normal operating conditions separately from short peaks. Include units with every numerical value. A drawing with the available space, orientation and access for removal gives the supplier a practical starting point.

Add the quantity for each position and identify any accessories that belong with it. Where a particular drawing, material certificate or calibration record is required, mention it with the line item. Submit the list with the destination country so that the response can cover the requested configuration and delivery.