DTL C
Electrical test & measurement

Current & voltage transducers
The i6000s Flex is a flexible AC current probe based on the Rogowski principle, designed for measuring high AC currents. Its flexible, lightweight measuring head is wrapped around the conductor and can be installed in cramped spaces and hard-to-reach areas.
A Rogowski coil is an air-cored coil wound around the conductor; it produces a voltage proportional to the rate of change of the current, which the probe's electronics integrate into an output signal proportional to the current. The i6000s Flex outputs a voltage signal that is compatible with oscilloscopes, recorders and data loggers.
Three ranges can be selected with a switch: 60 A, 600 A and 6000 A AC rms, with output sensitivities of 50 mV, 5 mV and 0.5 mV per ampere. Accuracy at 25 °C is ±1 % of range from 45 to 65 Hz, and linearity from 10 % to 100 % of range is ±0.2 % of reading. The frequency range is 10 Hz to 50 kHz (−3 dB). The working voltage is 600 V AC rms or DC, and the instrument connected must have a load impedance of at least 100 kΩ.
Two head lengths are available: 610 mm (24 in) and 915 mm (36 in). The head cable has a diameter of 14.3 mm and a bend radius of 38.1 mm, and the head weighs 180 g. A 2 m cable connects the head to the electronics, and the output is a 0.5 m cable terminated with a safety BNC connector; a 4 mm safety plug adaptor is available. The head operates from −20 to +90 °C and the electronics from −20 to +85 °C, at 15 to 85 % humidity, non-condensing.
The probe is designed to BS EN 61010-1, BS EN 61010-2-032 and BS EN 61010-031, and EMC to BS EN 61326. Listed articles include the 24-inch version and the RR6030 and RR6030/36 variants.
Typical uses are measurements with portable oscilloscopes, power quality analysis and electrical testing, wherever high AC currents must be measured on conductors that are difficult to reach.
Please state the required head length and the instrument to which the probe will be connected, so that the correct output adapter can be included with your quotation.
i6000s Flex-24, I6000S Flex-24, RR6030, RR6030/36
Match the instrument to the circuit and measurement category. Include the wiring arrangement and ratios of any external current or voltage transformers.
For an installed instrument, copy the complete ordering code and its suffixes from the nameplate. Identify the part that will remain in service and the part that needs to be replaced. A photograph of the connection face or terminal markings helps resolve differences between versions that share the same housing.
For a new installation, describe what the instrument must measure or control and where the reading will be used. Give the normal operating conditions separately from short peaks. Include units with every numerical value. A drawing with the available space, orientation and access for removal gives the supplier a practical starting point.
Add the quantity for each position and identify any accessories that belong with it. Where a particular drawing, material certificate or calibration record is required, mention it with the line item. Submit the list with the destination country so that the response can cover the requested configuration and delivery.