CHINO Corporation NR
CHINO Corporation

NR mineral-insulated sheathed Pt100 resistance thermometer

RTD sensors

Manufacturer
CHINO Corporation
Model
NR
Instrument type
RTD sensors
Element
Pt100, JPt100
Tolerance class
JIS A or B, 3-wire
Measuring current
2 mA or 1 mA
Measuring temperature
−200 to 500 °C
Insulation
high-purity magnesium oxide (MgO)
Sheath material
SUS316
Sheath outer diameter
φ3.2, 4.8, 6.4, 8 mm
Sheath length
300 to 2000 mm
Tip
closed type
Bending radius
at least 2 to 3 times the sheath diameter; not within 100 mm of tip
Mounting brackets
not within 70 mm of the tip
Wiring
3-wire; 4-wire (1 pair) option
Termination
Z-type terminal box or headless with direct wires
Options
explosion-proof versions
Request a quote
Review enquiry

The NR is a sheathed resistance thermometer. Its platinum resistance element is inserted into a metal protective tube and sealed with high-purity magnesium oxide powder, an inorganic insulation. The result is a sensor with good insulation and airtightness that resists vibration and mechanical shock and whose sheath can be bent on site.

A resistance thermometer measures temperature through the change of electrical resistance of platinum, which increases with temperature in a well-defined way. In the NR there are no air layers inside the protective tube, so heat reaches the element quickly; the response is stated to be two to three times better than that of general types. Because the sheath can be bent, L-shaped configurations can be made without elbows.

The element is Pt100 or JPt100, with tolerance to JIS class A or B in 3-wire connection, and a measuring current of 2 mA or 1 mA. The measuring range is −200 °C to 500 °C. The protective tube is SUS316 with an outer diameter of φ3.2, 4.8, 6.4 or 8 mm and a length from 300 to 2000 mm; the tip is closed. The standard construction uses a two-pair protective tube with three lead wires, and a 4-wire version with one pair is available.

When bending, the radius must be at least two to three times the tube diameter, and the 100 mm closest to the tip must stay straight. Mounting brackets cannot be installed within 70 mm of the tip insulation. Various mounting brackets are offered.

The standard termination is a Z-type hatch-shaped terminal box for easy connection; headless types with direct wire connection are also available, as are explosion-proof versions. Variants NRZR and NRHS1-0 with different tip or connection designs are listed.

Listed sectors are semiconductor, electronics, steel, aircraft, decarbonisation, chemical, pharmaceuticals, food and IoT.

For a quotation, please state the element and tolerance class, wiring, sheath diameter and length, termination and mounting fitting, and whether an explosion-proof version is required.

Industry applications

Selection notes

Match the sensing element, insertion length and wiring to the existing input. Separate the process temperature from the ambient temperature at the connection head.

For an installed instrument, copy the complete ordering code and its suffixes from the nameplate. Identify the part that will remain in service and the part that needs to be replaced. A photograph of the connection face or terminal markings helps resolve differences between versions that share the same housing.

For a new installation, describe what the instrument must measure or control and where the reading will be used. Give the normal operating conditions separately from short peaks. Include units with every numerical value. A drawing with the available space, orientation and access for removal gives the supplier a practical starting point.

Add the quantity for each position and identify any accessories that belong with it. Where a particular drawing, material certificate or calibration record is required, mention it with the line item. Submit the list with the destination country so that the response can cover the requested configuration and delivery.