EDT 101
Temperature

RTD sensors
The NR800-6 and NR800-7 are intrinsically safe, explosion-proof platinum-cobalt sheathed resistance thermometers for cryogenic temperature measurement. They measure down to 4 K (NR800-6) or 15 K (NR800-7), up to 375 K, and are designed so that, when combined with a barrier, they can be used in Class 0 hazardous locations with liquid hydrogen.
The sensing element is a platinum-cobalt resistor with a nominal value of 100 Ω ±0.15 Ω at 0 °C, operated with a measuring current of 1 mA. Its sensitivity is at least 0.09 Ω/K at 12 K and up to 0.40 Ω/K. The tolerance is ±0.5 K.
To pass the intrinsically safe explosion-proof equipment inspection, the sheathed sensor is combined with an insulation barrier. Insulation is improved by a φ1.6 mm sheathed wire between the sensing part and the sleeve. The sensing part sits in a capsule-type protective tube of oxygen-free copper with gold plating, φ2 × 30 mm, and the sheath is made of SUS316 with a diameter of 1.6 mm. The sheath wire can be installed in a bent state.
The explosion class and ignition level are IIC T4 Ga. For the intrinsically safe circuit, the permissible voltage is 12.4 V, the permissible current 42 mA and the allowable power 69.3 mW. The type inspection approval number is TC22088X.
The length of protective tube and sheath is specified between 50 mm and 100,000 mm. The lead is a four-wire vinyl-coated copper conductor, available in lengths from 500 mm to 600,000 mm.
Two versions are available: NR800-6 for 4 K to 375 K and NR800-7 for 15 K to 375 K. Applications named for these thermometers include pharmaceutical, environmental and medical measurement and monitoring, and GDP temperature control. When you enquire, please state the required version, the measuring range, the protective tube and sheath length, the lead length and the barrier or hazardous area conditions of the installation.
NR800-6, NR800-7
Match the sensing element, insertion length and wiring to the existing input. Separate the process temperature from the ambient temperature at the connection head.
For an installed instrument, copy the complete ordering code and its suffixes from the nameplate. Identify the part that will remain in service and the part that needs to be replaced. A photograph of the connection face or terminal markings helps resolve differences between versions that share the same housing.
For a new installation, describe what the instrument must measure or control and where the reading will be used. Give the normal operating conditions separately from short peaks. Include units with every numerical value. A drawing with the available space, orientation and access for removal gives the supplier a practical starting point.
Add the quantity for each position and identify any accessories that belong with it. Where a particular drawing, material certificate or calibration record is required, mention it with the line item. Submit the list with the destination country so that the response can cover the requested configuration and delivery.