Type 8189
Level

Radar level transmitters
NRM is a two-wire non-contact radar level transmitter for continuous measurement of level, distance or volume of liquids in tanks and vessels. It works with pulse burst radar at 24 or 25 GHz: the electronics send nanosecond microwave pulses through the antenna, receive the echo from the product surface and calculate the distance from the time of flight. Since the propagation speed of microwaves does not depend on process temperature and pressure, the measurement is not affected by changes in these conditions, and nothing touches the medium.
The measuring range is from 0.2 to 23 m, with an accuracy of ±3 mm. The high K-band frequency allows small antennas and a well-focused beam: depending on the horn or parabolic antenna, the beam angle is between 6° and 27°, which helps to avoid internal obstacles in the tank. The dead zone close to the antenna is between 135 and 354 mm. Media that tend to vaporise or form gas or fog layers can be measured, and vacuum applications are possible.
The transmitter is available in two versions: the compact NRM-4 and the integrated NRM-7. It is a loop-powered two-wire instrument with a 4–20 mA HART output and a supply of 20 to 36 V DC, or 20 to 30 V DC for Ex versions. Configuration is done via HART, for example with a USB HART modem and PC software, and a separate NRM-300P display unit can show the reading locally.
Maximum process pressure is 25 bar and maximum process temperature 180 °C; a high-temperature version is offered. Wetted parts and housings are available in PP, aluminium, stainless steel 1.4571 and PBT. Process connections include BSP and NPT threads, Tri-Clamp for hygienic applications, flanges and pipe couplings. Protection is IP67 or IP68. Ex versions are approved to ATEX and IECEx, and there are also marine and sanitary versions.
Typical media are petroleum products, crude oil, solvents such as acetone and ethanol, chemicals and water. Applications are found in food, power, water and wastewater, pharmaceutical and chemical plants as well as on ships. Please indicate tank height, medium, pressure, temperature, connection and whether an Ex approval is required.
NRM-4, NRM-7, NRM-300P, NRM-300P display unit, NUS-NTB-NRM-SW, NRM-4BSR9T100, HARTCOMM
Identify the vessel height, nozzle position and surface conditions. A point alarm and a continuous level reading require different outputs.
For an installed instrument, copy the complete ordering code and its suffixes from the nameplate. Identify the part that will remain in service and the part that needs to be replaced. A photograph of the connection face or terminal markings helps resolve differences between versions that share the same housing.
For a new installation, describe what the instrument must measure or control and where the reading will be used. Give the normal operating conditions separately from short peaks. Include units with every numerical value. A drawing with the available space, orientation and access for removal gives the supplier a practical starting point.
Add the quantity for each position and identify any accessories that belong with it. Where a particular drawing, material certificate or calibration record is required, mention it with the line item. Submit the list with the destination country so that the response can cover the requested configuration and delivery.