Multifunction process calibrators
The OZGU-370SE is an ozone transfer standard: a calibration unit for the dynamic and continuous production of zero and span gas for ozone analysers. It is used for quality control of ozone monitors in measuring stations and in the laboratory, and for preparing zero and span gas for multipoint O3 calibration.
Ozone is generated by UV irradiation from a mercury lamp whose intensity is controlled and which is housed in an oven. The concentration actually produced is measured by an internal type-approved measuring cell. This measured value serves as the set point for the second ozone analyser – the instrument being calibrated – so that the reading of the analyser under test can be compared directly with the transfer standard.
The unit performs multipoint calibration according to EN 14625. Twenty span points can be selected in the range 0 to 1000 ppb, and eight cycles are available. Calibration sequences can be programmed and started by a timer. Operation is through touch-screen displays.
For communication with a data acquisition system or a PC, the OZGU-370SE provides RS-232 and Ethernet interfaces. An internal data logger is available as an option. The unit is powered from 230 V AC, 50 Hz.
The device measures 430 mm wide, 550 mm deep and 221 mm high and weighs 17 kg. A portable case is available as an option, as are a special UV lamp (CPQ-8153, L = 117 mm) and an external zero air supply. The transfer standard holds TÜV approval No. 936/21204643/A.
By combining ozone generation and a reference measurement in one instrument, the OZGU-370SE lets the operator check and adjust ozone monitors at several concentration levels in one sequence, either on site at the station or in the laboratory.
When requesting a quotation, please state the required options (data logger, portable case, external zero air supply, spare UV lamp) and the number of units.
OZGU-370SE
Identify the instrument under test and the points to reproduce. State the required uncertainty, reference conditions and certificate format.
For an installed instrument, copy the complete ordering code and its suffixes from the nameplate. Identify the part that will remain in service and the part that needs to be replaced. A photograph of the connection face or terminal markings helps resolve differences between versions that share the same housing.
For a new installation, describe what the instrument must measure or control and where the reading will be used. Give the normal operating conditions separately from short peaks. Include units with every numerical value. A drawing with the available space, orientation and access for removal gives the supplier a practical starting point.
Add the quantity for each position and identify any accessories that belong with it. Where a particular drawing, material certificate or calibration record is required, mention it with the line item. Submit the list with the destination country so that the response can cover the requested configuration and delivery.