CHINO Corporation SC / NC intrinsically safe
CHINO Corporation

SC / NC intrinsically safe explosion-proof thermocouples

Thermocouples

Manufacturer
CHINO Corporation
Model
SC / NC intrinsically safe
Instrument type
Thermocouples
Series
SC (sheathed), NC (solid-pack)
Thermocouple types
R, N, K, E, J, T, etc.
Explosion protection
intrinsically safe (ia)
Explosion class and temperature class
IIC T5
Permissible voltage of IS circuit
10 V
Permissible current of IS circuit
400 mA
Type inspection approval
No. TC19165
Protection tube outer diameter
φ1.0 to 22 mm
Length
50 to 50,000 mm
Barriers
Zener barrier or isolating barrier
Request a quote
Review enquiry

The SC and NC series are thermocouples of intrinsically safe explosion-proof construction (ia) for temperature measurement in hazardous areas where flammable gases or vapours may be present. The SC series covers sheathed thermocouples and the NC series solid-pack sensors; both have passed the intrinsically safe explosion-proof equipment inspection in combination with Zener barriers or isolating barriers.

A thermocouple generates a small voltage from the temperature difference between its measuring junction and the connection end. In an intrinsically safe system the energy that can reach the sensor in the hazardous area is limited by a barrier in the safe area so that no spark or hot surface can ignite the surrounding atmosphere. For the SC and NC series the permissible voltage of the intrinsically safe circuit is 10 V and the permissible current 400 mA.

The sensors are classified for explosion class and ignition level IIC T5 of the target gas or vapour. The type inspection approval number is TC19165, issued after inspection by the Technology Association of Industrial Safety according to Japanese standards.

Thermocouple types R, N, K, E, J and T, among others, are available. Protection tubes are made with outer diameters from φ1.0 mm to 22 mm and in lengths from 50 mm up to 50,000 mm, so both small probes and long sensors can be built in explosion-proof form.

Two barrier arrangements are possible. With a Zener barrier, DIN-rail mounting and the intrinsic safety earth connection are made at the same time. With an isolating barrier, no earthing work is required.

Listed sectors are chemicals, pharmaceuticals, food, semiconductor, electronics, steel, aircraft and decarbonisation.

For a quotation, please state the thermocouple type, the protection tube diameter and length, the sheathed or solid-pack design, the gas group of the area and the barrier type to be used.

Industry applications

Selection notes

Match the sensing element, insertion length and wiring to the existing input. Separate the process temperature from the ambient temperature at the connection head.

For an installed instrument, copy the complete ordering code and its suffixes from the nameplate. Identify the part that will remain in service and the part that needs to be replaced. A photograph of the connection face or terminal markings helps resolve differences between versions that share the same housing.

For a new installation, describe what the instrument must measure or control and where the reading will be used. Give the normal operating conditions separately from short peaks. Include units with every numerical value. A drawing with the available space, orientation and access for removal gives the supplier a practical starting point.

Add the quantity for each position and identify any accessories that belong with it. Where a particular drawing, material certificate or calibration record is required, mention it with the line item. Submit the list with the destination country so that the response can cover the requested configuration and delivery.

Related instruments