CHINO Corporation SC
CHINO Corporation

SC mineral-insulated sheathed thermocouple, φ0.1 to φ8.0 mm

Thermocouples

Manufacturer
CHINO Corporation
Model
SC
Instrument type
Thermocouples
Thermocouple types
K, J, E, T, N (R on request)
Temperature range
−200 to 1050 °C (depending on type and design)
Tolerance class
JIS class 1 (0.4 grade), class 2 (0.75 grade)
Insulation
high-purity magnesium oxide powder
Sheath outer diameter
φ0.1, 0.15, 0.3, 0.5, 1.0, 1.6, 3.2, 4.8, 6.4, 8.0 mm
Sheath material
K: SUS316, SUS310S, SUS347, NCF600 (Inconel); J, E, T: SUS316; N: NCF600
Sheath wall thickness
at least 10 % of sheath outer diameter
Thermoelement wire diameter
at least 15 % of sheath outer diameter
Standard lengths
300, 500, 750, 1000, 1500, 2000 mm
Maximum length
from 2 m (φ0.1, φ0.15) up to 200 m (φ1.0)
Bending radius
at least 2 to 3 times the sheath outer diameter
Terminations
terminal box or headless with direct compensating cable
Tip shapes
three types
High-temperature versions
Alloy B, Alloy X
Request a quote
Review enquiry

The SC is a sheathed (mineral-insulated) thermocouple: the thermocouple wires are placed inside a metal protective tube and embedded in high-purity magnesium oxide powder, an inorganic insulating material. This sealed design gives good insulation and airtightness and protects the thermoelements from hostile atmospheres, while the sheath can still be bent to follow the installation.

A thermocouple produces a small voltage that depends on the temperature difference between the measuring junction at the tip and the reference junction at the connection. Because the measuring junction of a sheathed design has a very small heat capacity, particularly in the ultra-thin sizes, the sensor responds quickly to temperature changes.

The SC range covers measurements from −200 °C to 1050 °C, depending on thermocouple type and construction. Types K, J, E, T and N are available, and type R can also be manufactured. Tolerances correspond to JIS class 1 (0.4 grade) or class 2 (0.75 grade). Versions with Alloy B and Alloy X sheaths are offered for higher temperatures.

Sheath outer diameters run from an ultra-thin φ0.1 mm through φ0.15, 0.3, 0.5, 1.0, 1.6, 3.2, 4.8 and 6.4 mm up to φ8.0 mm. The wall thickness of the sheath is at least 10 % and the wire diameter at least 15 % of the sheath outer diameter. Sheath material depends on the type: for type K, SUS316, SUS310S, SUS347 or NCF600 (Inconel); for J, E and T, SUS316; for N, NCF600. Standard lengths are 300, 500, 750, 1000, 1500 and 2000 mm. Maximum length depends on diameter, for example 2 m for φ0.1 and φ0.15 mm, 200 m for φ1.0 mm, 100 m for φ1.6 mm, 50 m for φ3.2 and φ4.8 mm and 15 m for φ8.0 mm. The minimum bending radius is two to three times the sheath diameter. The design is stated to have good resistance to vibration from earthquakes and to corrosion.

The sensor can be supplied with a terminal box or as a headless type with compensating cable connected directly. Three tip shapes and a range of mounting brackets are available.

Listed applications are temperature measurement in narrow spaces or on small objects, heat treatment furnaces and chemical plants.

For a quotation, please state the thermocouple type, tolerance class, sheath diameter, material and length, tip shape, termination and mounting fitting.

Industry applications

Selection notes

Match the sensing element, insertion length and wiring to the existing input. Separate the process temperature from the ambient temperature at the connection head.

For an installed instrument, copy the complete ordering code and its suffixes from the nameplate. Identify the part that will remain in service and the part that needs to be replaced. A photograph of the connection face or terminal markings helps resolve differences between versions that share the same housing.

For a new installation, describe what the instrument must measure or control and where the reading will be used. Give the normal operating conditions separately from short peaks. Include units with every numerical value. A drawing with the available space, orientation and access for removal gives the supplier a practical starting point.

Add the quantity for each position and identify any accessories that belong with it. Where a particular drawing, material certificate or calibration record is required, mention it with the line item. Submit the list with the destination country so that the response can cover the requested configuration and delivery.