HORIBA SEC-Z700S
HORIBA

SEC-Z700S pressure-insensitive thermal mass flow controller

Thermal mass meters & flow controllers

Manufacturer
HORIBA
Model
SEC-Z700S
Instrument type
Thermal mass meters & flow controllers
Type
Pressure-insensitive thermal mass flow controller
Full-scale flow (N2)
30 SCCM to 50 SLM
Accuracy
≤ ±1.0 % S.P. (≥ 5 % flow); ≤ ±0.05 % F.S. (< 5 %)
Linearity
≤ ±0.5 % F.S.
Repeatability
≤ ±0.15 % S.P.
Control range
0.5–100 % F.S.
Response time
450 ±30 ms; tunable 300–1000 ms
Zero stability
≤ ±0.3 % F.S./year
Pressure
Supply ≤ 450 kPa (G); differential 50–100 to 400 kPa (D); proof 1 MPa (G)
On-board pressure / temperature
±3.5 kPa (0–700 kPa abs); ±2 °C (15–60 °C)
Leak integrity
≤ 5 × 10⁻¹² Pa·m³/s (He)
Wetted materials
SUS316L, Ni alloy, PFA
Fittings
1/4" VCR equivalent; 1.125" C-seal or W-seal
Communication
DeviceNet (SEC-Z714SMG, Z724SMG); EtherCAT (SEC-Z717SMG, Z727SMG)
Power / weight
24 VDC, 7.5 VA; 1.1–1.3 kg
Request a quote
Review enquiry

The SEC-Z700S series is a pressure-insensitive mass flow controller for stable gas flow control that is not affected by pressure fluctuations. It integrates thermal flow sensors with on-board pressure and temperature measurement, which simplifies gas delivery systems: in multi-chamber gas lines, separate line regulators and pressure sensors are no longer needed.

Gas flow is measured with thermal sensors and controlled by an integrated valve, available as normally closed or normally open. The pressure-insensitive function keeps the flow rate stable regardless of upstream or downstream pressure changes. A multi-gas / multi-range function is provided.

Full-scale flow rates from 30 SCCM to 50 SLM (N2 equivalent) are available. Flow accuracy is ≤ ±1.0 % of set point for flows of 5 % and above, and ≤ ±0.05 % of full scale below 5 %. Linearity is ≤ ±0.5 % F.S., repeatability ≤ ±0.15 % S.P. and zero stability ≤ ±0.3 % F.S. per year. The control range is 0.5 to 100 % of full scale. Response time is 450 ±30 ms and can be tuned from 300 ms to 1000 ms. The on-board sensors measure pressure with ±3.5 kPa accuracy over 0–700 kPa (abs) and temperature with ±2 °C over 15–60 °C.

The supply pressure must not exceed 450 kPa (G); the operating differential pressure is up to 400 kPa, with a minimum of 50–100 kPa depending on range, and proof pressure is 1 MPa (G). Operating temperature is 15 to 60 °C, non-condensing.

Communication is via DeviceNet (SEC-Z714SMG, SEC-Z724SMG) or EtherCAT (SEC-Z717SMG, SEC-Z727SMG), with M12 5-pin, Ethernet and RS-485 connections. Power is 24 VDC, 7.5 VA.

The ultra-clean flow path uses SUS316L, nickel alloy and PFA, with a helium leak integrity of ≤ 5 × 10⁻¹² Pa·m³/s. Connections are 1/4" VCR-equivalent fittings or 1.125" C-seal and W-seal surface mounts; the body is 124 mm fitting-to-fitting (92 mm port-to-port) and weighs 1.1–1.3 kg. Mounting orientation is free, and a warm-up of at least 60 minutes is required. The device follows SEMI E56, E69 and F64 and conforms to RoHS, REACH and EMC.

Applications named include semiconductor manufacturing processes, research and testing laboratories and automotive manufacturing.

Please send the gas, full-scale flow, pressure conditions, fitting type, valve type and fieldbus with your enquiry.

Part numbers · 4

SEC-Z714SMG, SEC-Z717SMG, SEC-Z724SMG, SEC-Z727SMG

Industry applications

Selection notes

State the minimum, normal and peak flow with units. Pipe bore, fluid properties and the available straight run affect the choice of measuring principle.

For an installed instrument, copy the complete ordering code and its suffixes from the nameplate. Identify the part that will remain in service and the part that needs to be replaced. A photograph of the connection face or terminal markings helps resolve differences between versions that share the same housing.

For a new installation, describe what the instrument must measure or control and where the reading will be used. Give the normal operating conditions separately from short peaks. Include units with every numerical value. A drawing with the available space, orientation and access for removal gives the supplier a practical starting point.

Add the quantity for each position and identify any accessories that belong with it. Where a particular drawing, material certificate or calibration record is required, mention it with the line item. Submit the list with the destination country so that the response can cover the requested configuration and delivery.