ifm electronic SV
ifm electronic

SV vortex flow sensor with integrated temperature measurement

Vortex flowmeters

Manufacturer
ifm electronic
Model
SV
Instrument type
Vortex flowmeters
Measuring principle
vortex, piezoceramic detection, integrated temperature measurement
Media
water-based media, ultrapure water, glycol solutions, cooling water
Nominal width
DN6, DN8, DN10, DN15, DN20, DN25
Process connection
G1/2, G3/4, G1, G1 1/4
Accuracy, water
< 1 % of range end value (Q < 50 %); < 2 % of measured value (Q > 50 %)
Accuracy, glycol
±5 % (2–6 cSt); ±10 % (6–15 cSt) of range end value
Repeatability
0.2 % of range end value
Medium temperature
−15 to 125 °C
Pressure rating
16 bar (up to 90 °C)
Bursting pressure
≥ 25 bar (up to 90 °C)
Outputs
4–20 mA, 2 × 4–20 mA, Pt1000, 100–1000 Hz, IO-Link
Wetted materials
PPS, PPSU, FKM or EPDM seals
Housing
stainless steel or all-plastic
Protection
IP65 / IP67
Request a quote
Review enquiry

The SV series are compact vortex flow sensors that measure the volumetric flow of water-based media, ultrapure water and glycol solutions and, at the same time, the temperature of the medium. Both values are available from one inline device, and the output configuration is chosen with the article.

The sensor works on the vortex principle. The flowing medium generates alternating vortices behind a bluff body placed in the flow; a piezoceramic sensor detects these pressure fluctuations, and the vortex frequency is converted into the flow rate. The measuring tube contains no rotating parts. Temperature is measured by an integrated element and made available as a separate signal.

The series covers nominal widths from DN6 to DN25 with G1/2, G3/4, G1 and G1 1/4 process connections. For water, the measuring error is below 1 % of the range end value in the lower half of the range and below 2 % of the measured value in the upper half; repeatability is 0.2 % of the range end value. For glycol media the stated accuracy is ±5 % at a viscosity of 2 to 6 cSt and ±10 % at 6 to 15 cSt. Medium temperature may be −15 to 125 °C, the pressure rating is 16 bar and the bursting pressure at least 25 bar, both up to 90 °C.

Depending on the article, the sensor provides a 4–20 mA output, two 4–20 mA outputs, a Pt1000 temperature signal, a frequency output of 100 to 1000 Hz or IO-Link for digital transfer of measured values and parameters.

Wetted parts are made of PPS and PPSU with FKM or EPDM seals. Versions with a stainless steel housing and all-plastic versions made of hydrolysis-resistant plastic are available, with IP65 or IP67 protection. A drinking water approval is available, and a version with display can be ordered. Rotatable inline fittings allow quick integration into the pipework. The articles SV3051 to SV6151 differ in nominal width and output configuration.

Typical applications named for the series are cooling systems, drinking water supplies and indoor farming. Please send the medium, nominal width, flow range and required output with your enquiry so that the matching article can be selected.

Part numbers · 12

SV3051, SV3151, SV4051, SV4151, SV5051, SV5151, SV6051, SV6151, SV7051, SV7151, SV8051, SV8151

Industry applications

Selection notes

State the minimum, normal and peak flow with units. Pipe bore, fluid properties and the available straight run affect the choice of measuring principle.

For an installed instrument, copy the complete ordering code and its suffixes from the nameplate. Identify the part that will remain in service and the part that needs to be replaced. A photograph of the connection face or terminal markings helps resolve differences between versions that share the same housing.

For a new installation, describe what the instrument must measure or control and where the reading will be used. Give the normal operating conditions separately from short peaks. Include units with every numerical value. A drawing with the available space, orientation and access for removal gives the supplier a practical starting point.

Add the quantity for each position and identify any accessories that belong with it. Where a particular drawing, material certificate or calibration record is required, mention it with the line item. Submit the list with the destination country so that the response can cover the requested configuration and delivery.

Related instruments

Find equipment

DEVICEEU