Temperature

Pyrometers & thermal cameras
The UW series is a wireless non-contact infrared temperature sensor system. It consists of a remote infrared sensor head and a radio transmitter housed in a NEMA 4X (IP65) enclosure. It measures surface temperatures in confined, hard-to-reach or harsh industrial environments where running signal wires is impractical.
The sensor head detects the infrared radiation emitted by the target surface and converts it into a temperature reading. Emissivity is adjustable from 0.10 to 1.0, so the reading can be matched to the surface being measured, and the field of view is 6:1. Besides process temperature, the transmitter measures ambient temperature, signal strength and battery status, and broadcasts all values wirelessly to a receiver. Free software turns a PC into a multi-channel chart recorder or data logger that displays the readings in real time.
The measuring range is −18 to 538 °C (0 to 1000 °F). Accuracy at 22 °C ambient and an emissivity of 0.95 or greater is ±2 % of reading or 2.2 °C, whichever is greater; repeatability is ±1 % and the response time is 100 ms.
Data is transmitted wirelessly, with a USB receiver option for the PC. The transmitter is powered by a 3 V DC battery with a life of approximately 1.5 years. The sensor head is connected by a 1.8 m (6 ft) cable as standard, and total cable lengths of up to 15 m are possible with extension cables.
The enclosure is weather-resistant to NEMA 4X and IP65 and is designed for wall mounting with the included mounting nut. The system operates from 0 to 70 °C (transmitter −10 to 70 °C) at less than 95 % RH, non-condensing. EN 60825-1 and FDA 21 CFR 1040.10 are listed.
The UWIR-2-NEMA is the standard transmitter and the UWIR-2A-NEMA a companion variant. Accessories include a water cooling jacket for use up to 200 °C, an air purge collar, a laser sighting accessory, a mounting bracket, a DIN rail adaptor and extension cables of 15 and 25 ft.
Applications named for the sensor are industrial process monitoring, heating coil temperature monitoring, induction welder monitoring and harsh or confined environments. Please send the target temperature range, the distance to the target and the required accessories with your enquiry.
UWIR-2-NEMA, UWIR-2A-NEMA, UWTC-BATT-C, UWTC-CABLE, OS100-AP, OS100-LS, OS100-WC, OS100-CA15FT, OS100-CA25FT, OS100-DR, OS100-MB
Match the sensing element, insertion length and wiring to the existing input. Separate the process temperature from the ambient temperature at the connection head.
For an installed instrument, copy the complete ordering code and its suffixes from the nameplate. Identify the part that will remain in service and the part that needs to be replaced. A photograph of the connection face or terminal markings helps resolve differences between versions that share the same housing.
For a new installation, describe what the instrument must measure or control and where the reading will be used. Give the normal operating conditions separately from short peaks. Include units with every numerical value. A drawing with the available space, orientation and access for removal gives the supplier a practical starting point.
Add the quantity for each position and identify any accessories that belong with it. Where a particular drawing, material certificate or calibration record is required, mention it with the line item. Submit the list with the destination country so that the response can cover the requested configuration and delivery.