Honeywell Process Solutions VersaFlow Coriolis
Honeywell Process Solutions

VersaFlow Coriolis mass flow sensor, models 100, 200, 1000 and 6000

Coriolis flowmeters

Manufacturer
Honeywell Process Solutions
Model
VersaFlow Coriolis
Instrument type
Coriolis flowmeters
Measuring principle
Coriolis mass flow
Sensor models
Coriolis 100 (twin tube), 1000 (single straight tube), 200 (large diameter), 6000 (twin V-tube, extended temperature)
Measured variables
mass flow, density, temperature; calculated volume, concentration, velocity
Accuracy, liquid
±0.1 % of measured flow rate; ±0.05 % option (not in SIL mode)
Accuracy, gas
±0.35 % of measured flow rate + zero stability
Density
400–2500 kg/m³ (100/1000), 100–3000 kg/m³ (6000); ±2 kg/m³
Maximum flow
up to 170,000 kg/h (100), 420,000 kg/h (200), 550,000 kg/h (6000), 560,000 kg/h (1000)
Process temperature
−40 to +130 °C (100), −45 to +130 °C (200), −40 to +150 °C Ti / 0 to +100 °C SS (1000), −200 to +400 °C (6000)
Nominal pressure
up to 100 barg; 180 barg (200, Super Duplex); 200 barg (6000, Hastelloy)
Tube materials
stainless steel 316/316L, 304/304L, duplex UNS S31803, Super Duplex, Hastelloy C22, titanium, tantalum
Process connections
flanges EN 1092-1 / ASME / JIS; hygienic Tri-clamp, DIN 32676, ISO 2852, DIN 11864-2, SMS, IDF/ISS, RJT
Converter
TWC 9400 C compact or TWC 9400 F remote, up to 300 m
Outputs
current with HART, pulse, frequency, status, control input, Modbus RS-485
Supply
100–230 V AC, 22 VA; 24 V DC option, 12 W
Protection, approvals
IP67 / NEMA 4X; ATEX, IECEx, FM, CSA, PED, MID, OIML R117-1, SIL 2
Request a quote
Review enquiry

VersaFlow Coriolis mass flow sensors measure mass flow, density and temperature directly, and from these derive volume flow, concentration and velocity. They are used for liquids, gases and slurries, including viscous, shear-sensitive, cryogenic and highly corrosive fluids and products with entrained solids or gas. Four sensor models cover line sizes from DN04 to DN200.

The sensors work on the Coriolis principle: the measuring tubes are made to oscillate, and the mass flowing through them causes a phase shift between inlet and outlet that is proportional to mass flow, while the oscillation frequency gives density. Coriolis 100 is a general-purpose twin bent tube sensor; Coriolis 1000 has a single straight measuring tube for high-purity and chemical duties and is self-draining and easy to clean; Coriolis 200 is a large-diameter sensor for bulk and oil and gas applications; Coriolis 6000 is a twin V-tube sensor for extended temperature and cryogenic service. No straight inlet or outlet runs are needed, and the sensor can be mounted horizontally, vertically (preferably with uphill flow) or at an angle.

Liquid accuracy is ±0.1 % of measured flow as standard and ±0.05 % as an option for custody transfer; gas accuracy is ±0.35 % plus zero stability. Repeatability is better than 0.05 % (liquid) or 0.2 % (gas) plus zero stability, and zero stability can be as low as ±0.004 % of maximum flow. Density accuracy is ±2 kg/m³. Maximum flows range up to 560,000 kg/h depending on model, and process temperatures extend from −200 to +400 °C on the Coriolis 6000.

The sensors work with the TWC 9400 converter, mounted compact on the sensor (TWC 9400 C) or as a field, wall or rack unit up to 300 m away (TWC 9400 F), in Ex and non-Ex versions. Outputs include current with HART, pulse, frequency and status, a control input and Modbus over RS-485. The backlit graphic display is rotatable and operated by optical keys. Diagnostics follow NAMUR NE 107 and include entrained gas management and redundant calibration. Supply is 100–230 V AC (22 VA) or optionally 24 V DC (12 W).

Tube materials include 316/316L and 304/304L stainless steel, duplex and Super Duplex, Hastelloy C22, titanium and tantalum. Connections include EN 1092-1, ASME and JIS flanges and hygienic Tri-clamp, DIN 32676, ISO 2852, DIN 11864-2, SMS, IDF/ISS and RJT. Protection is IP67 / NEMA 4X. Approvals include ATEX, IECEx, FM, CSA with CRN, PED, MID, OIML R117-1, 3A, NACE MR0175 and SIL 2 to IEC 61508. Options include heating jackets, purge ports, polished surfaces and 3-, 5- or 10-point calibration.

Applications include custody transfer, bulk loading and unloading, pipeline and allocation metering, oil and gas, chemicals, food and beverage, pharmaceuticals, LNG and cryogenic duties.

Please send the medium, flow and pressure ranges, temperature, line size and required approvals.

Part numbers · 25

CM01, CM02, CM03, CM04, CM16, CM20, CM41, CM71, CM72, CM73, CM74, CM75, CM76, CM77, CM78, CM79, CM96, CM97, VersaFlow Coriolis 100, VersaFlow Coriolis 1000, TWC 9400 C, TWC 9400 F, TWC9400C, TWC9400F, CM93

Industry applications

Selection notes

State the minimum, normal and peak flow with units. Pipe bore, fluid properties and the available straight run affect the choice of measuring principle.

For an installed instrument, copy the complete ordering code and its suffixes from the nameplate. Identify the part that will remain in service and the part that needs to be replaced. A photograph of the connection face or terminal markings helps resolve differences between versions that share the same housing.

For a new installation, describe what the instrument must measure or control and where the reading will be used. Give the normal operating conditions separately from short peaks. Include units with every numerical value. A drawing with the available space, orientation and access for removal gives the supplier a practical starting point.

Add the quantity for each position and identify any accessories that belong with it. Where a particular drawing, material certificate or calibration record is required, mention it with the line item. Submit the list with the destination country so that the response can cover the requested configuration and delivery.