HORIBA X-52
HORIBA

X-52 digital simulator for checking conductivity meters

Conductivity & resistivity sensors

Manufacturer
HORIBA
Model
X-52
Instrument type
Conductivity & resistivity sensors
Function
simulator for conductivity measuring instruments
Conductivity outputs
19.00, 1.900 S/m; 190.0, 19.00, 1.900 mS/m; 10.00, 1.000 S/m; 100.0, 10.00, 1.000 mS/m; 100.0, 10.00 µS/m; 0.000 mS/m
Conductivity accuracy
±0.1 % ±1 digit to ±0.2 % ±1 digit (depending on output)
Temperature outputs
0.0, 25.0, 30.0, 60.0, 100.0 °C
Temperature accuracy
±0.1 °C or ±0.2 °C
Connectors
BNC (conductivity); miniature jack (temperature)
Power supply
2 × AA alkaline batteries, approx. 20 h continuous
Dimensions
170 × 89 × 40 mm (H × W × D)
Housing
ABS and acrylic resin
Operating conditions
5 to 40 °C, 0–80 % RH non-condensing
Storage temperature
0 to 45 °C
Accessories
1 m cable, soft case, strap
Compliance
EN IEC 61326-1 class B, EN IEC 63000, FCC class A
Request a quote
Review enquiry

The X-52 is a digital simulator for conductivity measuring instruments. It provides theoretical equivalent inputs that replace the electrode during the periodic inspection of the instrument, so the meter itself can be checked independently of the sensor. It is used for checking the linearity and the temperature compensation function of conductivity meters in laboratory water quality analysis, industrial process monitoring and research and testing laboratories, and for electrode calibration and verification.

The simulator outputs equivalent resistance values corresponding to defined conductivities, as well as thermistor-equivalent resistance values corresponding to defined temperatures. By selecting different values, the user checks whether the connected instrument displays the correct reading across its range and whether its temperature compensation responds correctly.

Conductivity values are available in span and linear modes: 19.00 S/m, 1.900 S/m, 190.0 mS/m, 19.00 mS/m and 1.900 mS/m, and 10.00 S/m, 1.000 S/m, 100.0 mS/m, 10.00 mS/m, 1.000 mS/m, 100.0 µS/m and 10.00 µS/m, plus a zero value of 0.000 mS/m. Depending on the value, the stated accuracy ranges from ±0.1 % ±1 digit to ±0.2 % ±1 digit. Temperature values are 0.0, 25.0, 30.0, 60.0 and 100.0 °C with an accuracy of ±0.1 °C or ±0.2 °C.

Connection to the instrument is through a BNC connector for conductivity inspection and a miniature jack connector for temperature inspection; the BNC connector is also listed for pH, mV, ION and DO inspection. To ensure accuracy, the specified cable must be used, and when the DS-10, ES-10 series or D-24 are inspected, the BNC and miniature jack connectors must not be connected at the same time.

Power comes from two AA alkaline batteries, giving approximately 20 hours of continuous operation. The housing, of ABS and acrylic resin, measures 170 × 89 × 40 mm. The simulator is operated at 5 to 40 °C and up to 80 % relative humidity without condensation, and stored at 0 to 45 °C. It complies with EN IEC 61326-1 class B, EN IEC 63000 and FCC class A. Accessories are a 1 m cable, soft case and strap; the order number is GZ9100496982K.

Please state the instruments to be inspected when enquiring.

Part numbers · 1

GZ9100496982K

Industry applications

Selection notes

Record the sample temperature, expected concentration and how the probe is mounted. Identify the mating transmitter when replacing a sensor.

For an installed instrument, copy the complete ordering code and its suffixes from the nameplate. Identify the part that will remain in service and the part that needs to be replaced. A photograph of the connection face or terminal markings helps resolve differences between versions that share the same housing.

For a new installation, describe what the instrument must measure or control and where the reading will be used. Give the normal operating conditions separately from short peaks. Include units with every numerical value. A drawing with the available space, orientation and access for removal gives the supplier a practical starting point.

Add the quantity for each position and identify any accessories that belong with it. Where a particular drawing, material certificate or calibration record is required, mention it with the line item. Submit the list with the destination country so that the response can cover the requested configuration and delivery.